Development and optimization of organic acid analysis in tobacco with ion chromatography and suppressed conductivity detection

被引:51
作者
Qiu, JS [1 ]
Jin, XH [1 ]
机构
[1] US Tobacco Mfg Ltd Partnership, Nashville, TN 37203 USA
关键词
tobacco; conductivity detection; central composite face-centered design; organic acids;
D O I
10.1016/S0021-9673(02)00034-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
With the aid of a central composite face-centered design, an ion chromatographic method was developed and optimized for analyzing organic acids in tobacco. A Dionex-100 ion chromatograph with an ion suppressor and a conductivity detector, and a Bio-Rad Aminex HPX-87H column were employed. Only 13 analyses were required to optimize two factors: column temperature and eluent strength. Two sets of optimal conditions for separating nine acids were found: 1.8 mM HFBA eluent and 42 degreesC column temperature, and 0.8 mM HFBA eluent and 50 degreesC column temperature. The flow-rate was 0.6 ml min(-1) and the analysis time was 18 min or less. A sample preparation procedure included extraction of 2 g ground tobacco with 100 ml of 5 mM sulfuric acid solution for 3 h, filtration of the extract, and dilution of the filtrate 10-fold with deionized water. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:81 / 88
页数:8
相关论文