Propeller EPI in the other direction

被引:68
作者
Skare, Stefan [1 ]
Newbould, Rexford D. [1 ]
Clayton, Dave B. [1 ]
Bammer, Roland [1 ]
机构
[1] Stanford Univ, Dept Radiol, Lucas MRSI Ctr, Stanford, CA 94305 USA
关键词
MRI; EPI; PROPELLER; susceptibility; diffusion-weighted imaging;
D O I
10.1002/mrm.20890
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new propeller EPI pulse sequence with reduced sensitivity to field inhomogeneities is proposed. Image artifacts such as blurring due to Nyquist ghosting and susceptibility gradients are investigated and compared with those obtained in previous propeller EPI studies. The proposed propeller EPI sequence uses a readout that is played out along the short axis of the propeller blade, orthogonal to the readout used in previous propeller methods. In contrast to long-axis readout propeller EPI, this causes the echo spacing between two consecutive phase-encoding (PE) lines to decrease, which in turn increases the k-space velocity in this direction and hence the pseudo-bandwidth. Long- and short-axis propeller EPI, and standard single-shot EPI sequences were compared on phantoms and a healthy volunteer. Diffusion-weighted imaging (DWI) was also performed on the volunteer. Short-axis propeller EPI produced considerably fewer image artifacts compared to the other two sequences. Further, the oblique blades for the long-axis propeller EPI were also prone to one order of magnitude higher residual ghosting than the proposed short-axis propeller EPI.
引用
收藏
页码:1298 / 1307
页数:10
相关论文
共 22 条
[1]   A model-based method for retrospective correction of geometric distortions in diffusion-weighted EPI [J].
Andersson, JLR ;
Skare, S .
NEUROIMAGE, 2002, 16 (01) :177-199
[2]   How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging [J].
Andersson, JLR ;
Skare, S ;
Ashburner, J .
NEUROIMAGE, 2003, 20 (02) :870-888
[3]  
BOWTELL R, 1994, P 2 ANN M ISMRM SAN, P411
[4]   IMAGE-RECONSTRUCTION FOR ECHO PLANAR IMAGING WITH NONEQUIDISTANT KAPPA-SPACE SAMPLING [J].
BRUDER, H ;
FISCHER, H ;
REINFELDER, HE ;
SCHMITT, F .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (02) :311-323
[5]   Removal of EPI Nyquist ghost artifacts with two-dimensional phase correction [J].
Chen, NK ;
Wyrwicz, AM .
MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (06) :1247-1253
[6]  
CHUANG TC, 2005, P 13 ANN M ISMRM MIA, P9
[7]  
CHUANG TV, 2004, P 12 ANN M ISMRM KY, P535
[8]   PROPELLER MRI: Clinical testing of a novel technique for quantification and compensation of head motion [J].
Forbes, KPN ;
Pipe, JG ;
Bird, CR ;
Heiserman, JE .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (03) :215-222
[9]   Elimination of Nyquist ghosting caused by read-out to phase-encode gradient cross-terms in EPI [J].
Grieve, SM ;
Blamire, AM ;
Styles, P .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (02) :337-343
[10]   SELECTION OF A CONVOLUTION FUNCTION FOR FOURIER INVERSION USING GRIDDING [J].
JACKSON, JI ;
MEYER, CH ;
NISHIMURA, DG ;
MACOVSKI, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) :473-478