Load-Sharing Between Anterior and Posterior Elements in a Lumbar Motion Segment Implanted With an Artificial Disc

被引:188
作者
Dooris, Andrew P. [1 ,2 ]
Goel, Vijay K. [1 ,2 ]
Grosland, Nicole M. [1 ,2 ]
Gilbertson, Lars G. [3 ]
Wilder, David G. [1 ,2 ]
机构
[1] Univ Iowa, Dept Biomed Engn, Iowa Spine Res Ctr, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Orthopaed, Iowa City, IA USA
[3] Univ Pittsburgh, Dept Orthopaed Surg, Musculoskeletal Res Ctr, Pittsburgh, PA USA
关键词
artificial disc; biomechanics; facets; finite element model; lumbar spine;
D O I
10.1097/00007632-200103150-00004
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. A nonlinear three-dimensional finite element model of the osteoligamentous L3-L4 motion segment was used to predict changes in posterior element loads as a function of disc implantation and associated surgical procedures. Objectives. To evaluate the effects of disc implantation on the biomechanics of the posterior spinal elements (including the facet joints, pedicles, and lamina) and on the vertebral bodies. Summary of Background Data. Although several artificial disc designs have been used clinically, biomechanical data-particularly the change in loads in the posterior elements after disc implantation-are sparse. Methods. A previously validated intact finite element model was implanted with a ball-and-cup-type artificial disc model via an anterior approach. The implanted model predictions were compared with in vitro data. To study surgical variables, small and large windows were cut into the anulus, and the implant was placed anteriorly and posteriorly within the disc space. The anterior longitudinal ligament was also restored. Models were subjected to either 800 N axial compression force alone or to a combination of 10 N-m flexion-extension moment and 400 N axial preload. Implanted model predictions were compared with those of the intact model. Results. Facet loads were more sensitive to the anteroposterior location of the artificial disc than to the amount of anulus removed. Under 800 N axial compression, implanted models with an anteriorly placed artificial disc exhibited facet loads 2.5 times greater than loads observed with the intact model, whereas posteriorly implanted models predicted no facet loads in compression. Implanted models with a posteriorly placed disc exhibited greater flexibility than the intact and implanted models with anteriorly placed discs. Restoration of the anterior longitudinal ligament reduced pedicle stresses, facet loads, and extension rotation to nearly intact levels. Conclusions. The models suggest that, by altering placement of the artificial disc in the anteroposterior direction, a surgeon can modulate motion-segment flexural stiffness and posterior load-sharing, even though the specific disc replacement design has no inherent rotational stiffness.
引用
收藏
页码:E122 / E129
页数:8
相关论文
共 45 条
[1]   SENSORY AND AUTONOMIC INNERVATION OF THE FACET JOINT IN THE RAT LUMBAR SPINE [J].
AHMED, M ;
BJURHOLM, A ;
KREICBERGS, A ;
SCHULTZBERG, M .
SPINE, 1993, 18 (14) :2121-2126
[2]  
[Anonymous], MODERN CERAMIC ENG
[3]   MORPHOLOGICAL BASIS FOR BACK PAIN - THE DEMONSTRATION OF NERVE-FIBERS AND NEUROPEPTIDES IN THE LUMBAR FACET JOINT CAPSULE BUT NOT IN LIGAMENTUM-FLAVUM [J].
ASHTON, IK ;
ASHTON, BA ;
GIBSON, SJ ;
POLAK, JM ;
JAFFRAY, DC ;
EISENSTEIN, SM .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1992, 10 (01) :72-78
[4]   The artificial disc: Theory, design and materials [J].
Bao, QB ;
McCullen, GM ;
Higham, PA ;
Dumbleton, JH ;
Yuan, HA .
BIOMATERIALS, 1996, 17 (12) :1157-1167
[5]   SUBSTANCE-P INNERVATION OF LUMBAR SPINE FACET JOINTS [J].
BEAMAN, DN ;
GRAZIANO, GP ;
GLOVER, RA ;
WOJTYS, EM ;
CHANG, V .
SPINE, 1993, 18 (08) :1044-1049
[6]  
Boas R A, 1991, Neurosurg Clin N Am, V2, P807
[7]   THE USE OF DENSE ALUMINA ALUMINA CERAMIC COMBINATION IN TOTAL HIP-REPLACEMENT [J].
BOUTIN, P ;
CHRISTEL, P ;
DORLOT, JM ;
MEUNIER, A ;
DEROQUANCOURT, A ;
BLANQUAERT, D ;
HERMAN, S ;
SEDEL, L ;
WITVOET, J .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (12) :1203-1232
[8]   DISKS DEGENERATE BEFORE FACETS [J].
BUTLER, D ;
TRAFIMOW, JH ;
ANDERSSON, GBJ ;
MCNEILL, TW ;
HUCKMAN, MS .
SPINE, 1990, 15 (02) :111-113
[9]   BIOMECHANICS OF THE SB CHARITE LUMBAR INTERVERTEBRAL-DISK ENDOPROSTHESIS [J].
BUTTNERJANZ, K ;
SCHELLNACK, K ;
ZIPPEL, H .
INTERNATIONAL ORTHOPAEDICS, 1989, 13 (03) :173-176
[10]   NEURAL MECHANISMS OF LUMBAR PAIN [J].
CAVANAUGH, JM .
SPINE, 1995, 20 (16) :1804-1809