Effect of Working Fluid on Sub-2 nm Particle Detection with a Laminar Flow Ultrafine Condensation Particle Counter

被引:153
作者
Iida, Kenjiro [1 ]
Stolzenburg, Mark R. [1 ]
McMurry, Peter H. [1 ]
机构
[1] Univ Minnesota, Particle Technol Lab, Minneapolis, MN USA
关键词
DIFFERENTIAL MOBILITY ANALYZER; HETEROGENEOUS NUCLEATION; THERMAL-DECOMPOSITION; AEROSOL; SIZE; EFFICIENCY; DEPOSITION; TOXICITY; CLUSTERS; NUCLEI;
D O I
10.1080/02786820802488194
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The effect of working fluid on size-dependent activation efficiencies with the laminar flow ultrafine condensation particle counter described by Stolzenburg and McMurry (1991) was studied theoretically and experimentally. Criteria considered include tendency to avoid homogeneous nucleation within the condenser and toxicity. The working fluids that were identified have vapor pressures below that of butanol, so particles grow to smaller sizes and are more difficult to detect optically. Therefore we use a second, conventional CPC as a " booster" to grow particles to a detectable size. Experiments were performed to obtain the size-and material-dependent activation efficiencies for ethylene glycol, diethylene glycol, propylene glycol, oleic acid, and DOS. Using diethylene glycol and oleic acid, values of the 50% activation efficiency diameter, D-p50, for negatively charged particles generated by evaporating sodium chloride, ammonium sulfate, and silver were < 1.2 (< 0.8) nm, 1.4-1.5 (1.0-1.2) nm, and 1.9-2.0 (1.5) nm as mobility (mass) diameter, respectively. The stability of the UCPC for long-term operation using ethylene glycol and propylene glycol as working fluids was tested by monitoring the instrument's response to silver particles having size near Dp50 in an air stream at 40-45% relative humidity. The performance was steady (+/- 3%) for several days indicating the instrument performs stably during unattended operation for realistic atmospheric sampling conditions.
引用
收藏
页码:81 / 96
页数:16
相关论文
共 55 条
[1]   Particle size dependent response of aerosol counters [J].
Ankilov, A ;
Baklanov, A ;
Colhoun, M ;
Enderle, KH ;
Gras, J ;
Julanov, Y ;
Kaller, D ;
Lindner, A ;
Lushnikov, AA ;
Mavliev, R ;
McGovern, F ;
O'Connor, TC ;
Podzimek, J ;
Preining, O ;
Reischl, GP ;
Rudolf, R ;
Sem, GJ ;
Szymanski, WW ;
Vrtala, AE ;
Wagner, PE ;
Winklmayr, W ;
Zagaynov, V .
ATMOSPHERIC RESEARCH, 2002, 62 (3-4) :209-237
[2]   RESPONSE CHARACTERISTICS FOR 4 DIFFERENT CONDENSATION NUCLEUS COUNTERS TO PARTICLES IN THE 3-50 NM DIAMETER RANGE [J].
BARTZ, H ;
FISSAN, H ;
HELSPER, C ;
KOUSAKA, Y ;
OKUYAMA, K ;
FUKUSHIMA, N ;
KEADY, PB ;
KERRIGAN, S ;
FRUIN, SA ;
MCMURRY, PH ;
PUI, DYH ;
STOLZENBURG, MR .
JOURNAL OF AEROSOL SCIENCE, 1985, 16 (05) :443-456
[3]   Relative toxicity of di(2-ethylhexyl) sebacate and related compounds in an in vivo hamster bioassay [J].
Brain, JD ;
Blanchard, JD ;
Heyder, J ;
Wolfthal, SF ;
Beck, BD .
INHALATION TOXICOLOGY, 1996, 8 (06) :579-593
[4]   A high efficiency, high throughput unipolar aerosol charger for nanoparticles [J].
Chen, Da-Ren ;
Pui, David Y. H. .
JOURNAL OF NANOPARTICLE RESEARCH, 1999, 1 (01) :115-126
[5]   Design and evaluation of a nanometer aerosol differential mobility analyzer (Nano-DMA) [J].
Chen, DR ;
Pui, DYH ;
Hummes, D ;
Fissan, H ;
Quant, FR ;
Sem, GJ .
JOURNAL OF AEROSOL SCIENCE, 1998, 29 (5-6) :497-509
[6]   Differential mobility analysis of molecular ions and nanometer particles [J].
de la Mora, JF ;
de Juan, L ;
Eichler, T ;
Rosell, J .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1998, 17 (06) :328-339
[7]   TOXICITY OF 29 PLASTICIZERS TO HELA-CELLS IN THE MIT-24 SYSTEM [J].
EKWALL, B ;
NORDENSTEN, C ;
ALBANUS, L .
TOXICOLOGY, 1982, 24 (3-4) :199-210
[8]  
Friedlander SK., 2000, SMOKE DUST HAZE FUND
[9]  
Fuchs NA., 1970, HIGHLY DISPERSED AER
[10]   A condensation nucleus counter (CNC) sensitive to singly charged sub-nanometer particles [J].
Gamero-Castaño, M ;
de la Mora, JF .
JOURNAL OF AEROSOL SCIENCE, 2000, 31 (07) :757-772