Direct quantification of transmitral flow volume with dynamic 3-dimensional digital color Doppler: A validation study in an animal model

被引:24
作者
Rusk, RA
Li, XN
Mori, Y
Irvine, T
Jones, M
Zetts, AD
Kenny, A
Sahn, DJ
机构
[1] Oregon Hlth Sci Univ, Portland, OR 97201 USA
[2] ATL Ultrasound, Bothell, WA USA
[3] NHLBI, Lab Anim Med & Surg, Bethesda, MD 20892 USA
[4] Freeman Rd Hosp, Newcastle Upon Tyne NE7 7DN, Tyne & Wear, England
关键词
D O I
10.1067/mje.2002.116716
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Accurately quantifying transmitral flow volume is clinically important not only as a measure of cardiac output, but also as a value from which to subtract aortic flow, for determining the severity of mitral regurgitation. However, controversy exists over the accuracy of pulsed Doppler for mitral flow quantification because of the complexity of mitral flow geometry and dynamic changes in flow profile and flow area. To explore the feasibility of directly quantifying transmitral flow volume with a newly developed dynamic 3-dimensional digital color Doppler technique, this in vivo experimental study was conducted to validate the method. Eight open chest sheep were imaged with a multiplane transesophageal (TEE) probe placed on the heart for digital 3-dimensional gated acquisition of mitral inflow over a 180-degree acquisition. The digital velocity data were contour detected for flow area after computing the velocity vectors and flow profile perpendicular to a spherical 3-dimensional surface across the mitral annulus. Flow areas and actual velocities were then integrated in time and space and the resulting flow volumes were compared with those obtained by a reference electromagnetic flowmeter on the aorta for 26 steady hemodynamic states. The flow volumes correlated closely to the electromagnetic references (y = 0.87x + 2.49, r = 0.92, SEE = 1.9 mL per beat). Our study shows that transmitral flow volume can be accurately determined in vivo by this dynamic 3-dimensional digital color Doppler flow quantification method.
引用
收藏
页码:55 / 62
页数:8
相关论文
共 27 条
[1]   CALCULATION OF TRANSMITRAL FLOW BY DOPPLER ECHOCARDIOGRAPHY - A COMPARISON OF METHODS IN A CANINE MODEL [J].
ASCAH, KJ ;
STEWART, WJ ;
GILLAM, LD ;
TRIULZI, MO ;
NEWELL, JB ;
WEYMAN, AE .
AMERICAN HEART JOURNAL, 1989, 117 (02) :402-411
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   INFLUENCE OF SAMPLING SITE AND FLOW AREA ON CARDIAC-OUTPUT MEASUREMENTS BY DOPPLER ECHOCARDIOGRAPHY [J].
DITTMANN, H ;
VOELKER, W ;
KARSCH, KR ;
SEIPEL, L .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (04) :818-823
[4]   THE MITRAL-VALVE ORIFICE METHOD FOR NON-INVASIVE TWO-DIMENSIONAL ECHO DOPPLER DETERMINATIONS OF CARDIAC-OUTPUT [J].
FISHER, DC ;
SAHN, DJ ;
FRIEDMAN, MJ ;
LARSON, D ;
VALDESCRUZ, LM ;
HOROWITZ, S ;
GOLDBERG, SJ ;
ALLEN, HD .
CIRCULATION, 1983, 67 (04) :872-877
[5]   Automated assessment of mitral regurgitant volume and regurgitant fraction by a newly developed digital color Doppler velocity profile integration method [J].
Hozumi, T ;
Yoshida, K ;
Akasaka, T ;
Takagi, T ;
Yamamuro, A ;
Yagi, T ;
Yoshikawa, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 80 (10) :1325-1330
[6]   A digital 3-dimensional method for computing great artery flows: In vitro validation studies [J].
Irvine, T ;
Li, XN ;
Mori, Y ;
Wanitkun, S ;
Li, XK ;
Detmer, PR ;
Martin, RW ;
Pope, A ;
Schwartz, GA ;
Rusk, RA ;
Kenny, A ;
Sahn, DJ .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2000, 13 (09) :841-848
[7]  
Irvine T, 1998, CIRCULATION, V98, P715
[8]   2-DIMENSIONAL MITRAL FLOW VELOCITY PROFILES IN PIG MODELS USING EPICARDIAL DOPPLER-ECHOCARDIOGRAPHY [J].
KIM, WY ;
BISGAARD, T ;
NIELSEN, SL ;
POULSEN, JK ;
PEDERSEN, EM ;
HASENKAM, JM ;
YOGANATHAN, AP .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (02) :532-545
[9]   A new Doppler method for quantification of volumetric flow: In vivo validation using color Doppler [J].
Kim, WY ;
Poulsen, JK ;
Terp, K ;
Staalsen, NH .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 27 (01) :182-192
[10]   SKEWNESS OF INSTANTANEOUS MITRAL TRANSANNULAR FLOW-VELOCITY PROFILES IN NORMAL HUMANS [J].
KUPARI, M ;
JARVINEN, V ;
POUTANEN, VP ;
HEKALI, P .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 268 (03) :H1232-H1238