A 72-hour study to test the efficacy and safety of the "Mucus Slurper" in mechanically ventilated sheep

被引:31
作者
Bassi, Gianluigi Li
Curto, Francesco
Zanella, Alberto
Stylianou, Mario
Kolobow, Theodor
机构
[1] NHLBI, Sect Pulm & Cardiac Assist Devices, Pulm & Crit Cre Med Branch, NIH,PCCMB, Bethesda, MD 20892 USA
[2] NHLBI, Div Epidemiol & Clin Applicat, Off Biostat Res, NIH, Bethesda, MD 20892 USA
关键词
artificial ventilation; intratracheal intubation; airway obstruction; bacterial pneumonia; intensive care; evaluation studies;
D O I
10.1097/01.CCM.0000257332.62358.0E
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: In patients intubated with endotracheal tubes (ETTs), suctioning is routinely performed to remove mucus from within the ETT and trachea. The Mucus Slurper is a novel ETT with built-in suction ports arranged radially at its tip. We evaluated the safety and efficacy of the Mucus Slurper, compared with conventional tracheal suction, to prevent airway obstruction in sheep with the ETT and trachea oriented below horizontal. Design: Prospective randomized animal study. Setting. Animal research facility at the National Institutes of Health. Subjects: Twelve healthy sheep. Interventions: Sheep were randomized to be intubated with either the Mucus Slurper (study group) or a Hi-Lo Tracheal Tube (Mallinckrodt, St. Louis, MO) (control group) and mechanically ventilated for 72 hrs. In the study group, automatic, timed tracheal aspiration lasted 0.3 secs, was repeated every 2 mins, and was synchronized with the early expiratory phase. In the control group, tracheal suction was performed every 6 hrs or as required. Measurements and Main Results: In the control group, tracheal secretions accumulated progressively within the ETT and the trachea. In the study group, all mucus that reached the tip of the Mucus Slurper was aspirated, keeping the lumen of the ETT, and proximal trachea, free from secretions, In the study group, expiratory water trap protein concentration, a crude index to measure mucus drainage through the ETT, was consistently less than the control group (p < .001). At autopsy, no macroscopic injury to the tracheal mucosa was found in either group. In the study group, the respiratory circuit was less colonized than in the control group. There were no statistically significant differences between the two groups in bacterial colonization of the lungs/bronchi. Conclusions. The Mucus Slurper, combined with orientation of the trachea below horizontal, prevents accumulation of secretions within the lumen of the ETT and trachea, without need for conventional tracheal suctioning.
引用
收藏
页码:906 / 911
页数:6
相关论文
共 29 条
[1]  
Tablan Ofelia C, 2004, MMWR Recomm Rep, V53, P1
[2]  
[Anonymous], 1993, RESP CARE
[3]   Endotracheal tubes coated with antiseptics decrease bacterial colonization of the ventilator circuits, lungs, and endotracheal tube [J].
Berra, L ;
De Marchi, L ;
Yu, ZX ;
Laquerriere, P ;
Baccarelli, A ;
Kolobow, T .
ANESTHESIOLOGY, 2004, 100 (06) :1446-1456
[4]  
Berra L, 2006, INTENS CARE MED, V32, P888, DOI 10.1007/s00134-006-0125-6
[5]   Endotracheal tube intraluminal diameter narrowing after mechanical ventilation:: use of acoustic reflectometry [J].
Boqué, MC ;
Gualis, B ;
Sandiumenge, A ;
Rello, J .
INTENSIVE CARE MEDICINE, 2004, 30 (12) :2204-2209
[6]  
Branson R D, 1998, Respir Care Clin N Am, V4, P305
[7]   The effect of suction catheter insertion and tracheal stimulation in adults with severe brain injury [J].
Brucia, J ;
Rudy, E .
HEART & LUNG, 1996, 25 (04) :295-303
[8]   Closed system endotracheal suctioning maintains lung volume during volume-controlled mechanical ventilation [J].
Cereda, M ;
Villa, F ;
Colombo, E ;
Greco, G ;
Nacoti, M ;
Pesenti, A .
INTENSIVE CARE MEDICINE, 2001, 27 (04) :648-654
[9]  
Glass C, 1999, Am J Crit Care, V8, P93
[10]  
Hagler D A, 1994, Am J Crit Care, V3, P444