A clinical study of the human lens with a dynamic light scattering device

被引:24
作者
Datiles, MB
Ansari, RR
Reed, GF
机构
[1] NEI, Cataract Sect, Ophthalm Genet & Visual Funct Branch, NIH, Bethesda, MD 20892 USA
[2] NASA, Glenn Res Ctr, Cleveland, OH 44135 USA
[3] NEI, Biometry Branch, Div Epidemiol & Clin Res, Bethesda, MD 20892 USA
基金
美国国家卫生研究院;
关键词
cataract; lens; dynamic light scattering; corneal topography; clinical study;
D O I
10.1006/exer.2001.1106
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
A study was conducted to determine the potential usefulness and repeatability of a new dynamic light scattering (DLS) device for clinical studies of the human lens and early cataract. Studies using the cold cataract model showed this new device to be more sensitive than the Scheimpflug cataract imaging system in detecting the earliest cataractous changes. A miniaturized clinical DLS device developed by NASA using fiber optic probes was mounted on a Keratoscope (Optikon 2000), which has a 3-dimensional aiming system for accurate repeated sampling of the same area of the lens. A test/retest study was then conducted on the nuclear region of the lenses of 12 normal eyes. After a full, dilated eye examination, DLS data were obtained using the new device on the same eyes twice, 30-60 min apart. Particle size distributions and mean log particle size data were obtained. The mean percent differences between the larger and smaller of the test-retest pairs was 6.4 % (range 0.05-10.8 %); the between-test S.D. was 0.116. Actual numerical margin of error was +/-0.023. In addition, the mean coefficient of variation was 4.2 % (range 0.3-7.3 %), A useful clinical end point obtained from data produced by the device was the mean log particle size. These results suggest that the DLS will be useful in the detection and study of the beginning and earliest stages of cataract formation in humans.
引用
收藏
页码:93 / 102
页数:10
相关论文
共 25 条
[1]  
Ansari R R, 1999, Diabetes Technol Ther, V1, P159, DOI 10.1089/152091599317378
[2]   Measuring lens opacity: combining quasi-elastic light scattering with Scheimpflug imaging system [J].
Ansari, RR ;
Datiles, MB ;
King, JF ;
Leftwood, D .
OPHTHALMIC TECHNOLOGIES VIII, PROCEEDINGS OF, 1998, 3246 :35-42
[3]   A fiber optic probe for monitoring protein aggregation, nucleation and crystallization [J].
Ansari, RR ;
Suh, KI ;
Arabshahi, A ;
Wilson, WW ;
Bray, TL ;
DeLucas, LJ .
JOURNAL OF CRYSTAL GROWTH, 1996, 168 (1-4) :216-226
[4]   A new clinical instrument for the early detection of cataract using dynamic light scattering and corneal topography [J].
Ansari, RR ;
Datiles, MB ;
King, JF .
OPHTHALMIC TECHNOLOGIES X, 2000, 3908 :38-49
[5]   QUANTITATIVE DETECTION OF THE MOLECULAR-CHANGES ASSOCIATED WITH EARLY CATARACTOGENESIS IN THE LIVING HUMAN LENS USING QUASIELASTIC LIGHT-SCATTERING [J].
BENEDEK, GB ;
CHYLACK, LT ;
LIBONDI, T ;
MAGNANTE, P ;
PENNETT, M .
CURRENT EYE RESEARCH, 1987, 6 (12) :1421-1432
[6]   THEORY OF TRANSPARENCY OF EYE [J].
BENEDEK, GB .
APPLIED OPTICS, 1971, 10 (03) :459-&
[7]   CLINICAL PHOTON-CORRELATION SPECTROSCOPY EVALUATION OF HUMAN DIABETIC LENSES [J].
BURSELL, SE ;
BAKER, RS ;
WEISS, JN ;
HAUGHTON, JF ;
RAND, LI .
EXPERIMENTAL EYE RESEARCH, 1989, 49 (02) :241-258
[8]   LENS OPACITIES CLASSIFICATION SYSTEM-II (LOCS-II) [J].
CHYLACK, LT ;
LESKE, MC ;
MCCARTHY, D ;
KHU, P ;
KASHIWAGI, T ;
SPERDUTO, R .
ARCHIVES OF OPHTHALMOLOGY, 1989, 107 (07) :991-997
[9]  
DATILES M, 1988, OPHTHALMIC SURG LAS, V19, P664
[10]  
Datiles M., 1992, DUANES CLIN OPHTHALM, V1, P1