Comparison of different Immunoassays for CA 19-9

被引:26
作者
Stern, P
Friedecky, B
Bartos, V
Bezdickova, D
Vavrova, J
Uhrova, J
Rozprimova, L
Zima, T
Palicka, V
机构
[1] Gen Univ Hosp, Inst Clin Biochem, Prague, Czech Republic
[2] Charles Univ Prague, Fac Med 1, Prague, Czech Republic
[3] Inst Postgrad Educ Hlth Care Profess, Dept Clin Biochem, Prague, Czech Republic
[4] Univ Hosp, Inst Clin Biochem & Diagnost, Hradec Kralove, Czech Republic
[5] Univ Ostrava, Univ Hosp Ostrava, Dept Nucl Med, CZ-70852 Ostrava, Czech Republic
[6] Agilab Ltd, Prague, Czech Republic
关键词
CA; 19-9; immunoassays; repeatability; comparability; evidence-based medicine;
D O I
10.1515/CCLM.2001.205
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
We compared six routinely employed immunoassay kits: Architect i2000 and AxSYM, Abbott Laboratories; Elecsys 2010, Roche Diagnostics; ELSA, CIS-BioInternational; Immulite 1, Diagnostic Products Corporation; and IRMA-mat, Byk-Sangtec Diagnostica. Using all analytical systems, we measured identical groups of clinical samples completed with selected control samples. The repeatability of measurements (coefficient of variation) ranged from 2.1% (Elecsys 2010) to 6.7% (ELSA). The parameters of Passing-Bablok regression show significant systematic differences among analytical systems. Data from a Bland-Altman diagram suggest that these differences project onto other, still more significant individual differences among individual samples. Though the cut-off values differ between various systems, no similar clinical efficacy appears to be attained. The behavior of individual systems is quite different for identical control materials and does not necessarily duplicate the calibration for biological samples. The results of determining CA 19-9 cannot be extrapolated from one analytical technique to another, even in cases where the same monoclonal antibody is used. Standardization of CA 19-9 measurement systems is necessary to allow use of the results for the purposes of evidence-based medicine.
引用
收藏
页码:1278 / 1282
页数:5
相关论文
共 15 条
[1]  
*ABB LAB, 1996, INSTR MAN
[2]  
[Anonymous], EVIDENCE BASED MED P
[3]  
*ARCHITECT CA, 1998, 199 ARCHITECT CA ABB
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
Bruns DE, 2000, CLIN CHEM, V46, P893
[6]  
*BYKS DIAGN, 1999, INSTR MAN
[7]  
*CIS BIO INT, 2000, 16 ELSACA CIS BIO IN
[8]  
*DPC, 1997, IMMULITE GI MA PILKG
[9]   European symposium on clinical laboratory and in vitro diagnostics industry [J].
Dybkaer, R .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2001, 39 (05) :462-463
[10]   The science of systematic reviewing studies of diagnostic tests [J].
Oosterhuis, WP ;
Niessen, RWLM ;
Bossuyt, PMM .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2000, 38 (07) :577-588