Dose enhancement by a thin foil of high-Z material:: A Monte Carlo study

被引:31
作者
Li, XA
Chu, JCH
Chen, WM
Zusag, T
机构
[1] Rush Univ, Rush Presbyterian St Lukes Med Ctr, Dept Phys Med, Chicago, IL 60612 USA
[2] Rush Univ, Rush Presbyterian St Lukes Med Ctr, Dept Radiat Oncol, Chicago, IL 60612 USA
关键词
interface dosimetry; dose enhancement; EGS4 Monte Carlo; photon calculation;
D O I
10.1118/1.598619
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this work is to study the dose enhancement by a thin foil (thickness of 0.2-4 mm) of high-Z material in a water phantom, irradiated by high-energy photon beams. EGS4 Monte Carlo technique was used. Perturbations on the beam spectra due to the presence of the foils, and dose enhancement dependence of photon-beam quality, beam incident angle, atomic number (Z), the thickness and size of the foil, and the depth of the foil situated in the phantom were studied. Analysis of photon and secondary-electron spectra indicates that the dose enhancement near an inhomogeneity interface is primarily due to secondary electrons. A calculation for I-mm-thick planar lead foil in a water phantom shows that the dose enhancements at 0.25, 1, 2 and 3 mm away from the foil in the backward region were 58%, 37%, 24% and 17%, respectively, for a 15 MV beam. Calculations for a variety of planar foils and photon beams show that dose enhancement: (a) increases with Z; (b) decreases with decreasing foil thickness when the foils are thinner than a certain value (1 mm for lead foil for 15 MV); (c) decreases with decreasing incident photon-beam energies; (d) changes slightly for beam incident angles less than 45 degrees and more prominently for larger angles; (e) increases with size of foil; and (f) is almost independent of the depth at which the foil is situated when the foil is placed beyond the range of secondary electrons. The dose enhancement calculation is also performed for a cylindrically shaped lead foil irradiated by a four-field-box. The dose enhancement of 34%/13% was obtained at 0.25/2 mm away from the cylindrical outer interface for a 15 MV four-field-box. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)92107-0].
引用
收藏
页码:1245 / 1251
页数:7
相关论文
共 34 条
[1]  
Bielajew A., 1990, The Dosimetry of Ionizing Radiation, Volume, P427
[2]   THE MONTE-CARLO SIMULATION OF ION-CHAMBER RESPONSE TO CO-60 - RESOLUTION OF ANOMALIES ASSOCIATED WITH INTERFACES [J].
BIELAJEW, AF ;
ROGERS, DWO ;
NAHUM, AE .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (05) :419-427
[3]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[4]  
Bielajew AF, 1988, MONTE CARLO TRANSPOR, P115
[5]   DOSE ENHANCEMENT IN BUILDUP REGION BY LEAD, ALUMINUM, AND LUCITE ABSORBERS FOR 15 MVP PHOTON-BEAM [J].
CIESIELSKI, B ;
REINSTEIN, LE ;
WIELOPOLSKI, L ;
MEEK, A .
MEDICAL PHYSICS, 1989, 16 (04) :609-613
[6]   EFFECTS OF GOLD AND SILVER BACKINGS ON THE DOSE-RATE AROUND AN I-125 SEED [J].
CYGLER, J ;
SZANTO, J ;
SOUBRA, M ;
ROGERS, DWO .
MEDICAL PHYSICS, 1990, 17 (02) :172-178
[7]   BACKSCATTER DOSE PERTURBATION IN KILOVOLTAGE PHOTON BEAMS AT HIGH ATOMIC-NUMBER INTERFACES [J].
DAS, IJ ;
CHOPRA, KL .
MEDICAL PHYSICS, 1995, 22 (06) :767-773
[8]   Forward dose perturbation at high atomic number interfaces in kilovoltage x-ray beams [J].
Das, IJ .
MEDICAL PHYSICS, 1997, 24 (11) :1781-1787
[9]   VALIDITY OF TRANSITION-ZONE DOSIMETRY AT HIGH ATOMIC-NUMBER INTERFACES IN MEGAVOLTAGE PHOTON BEAMS [J].
DAS, IJ ;
KASE, KR ;
MEIGOONI, AS ;
KHAN, FM ;
WERNER, BL .
MEDICAL PHYSICS, 1990, 17 (01) :10-16
[10]   BACKSCATTER DOSE PERTURBATION AT HIGH ATOMIC-NUMBER INTERFACES IN MEGAVOLTAGE PHOTON BEAMS [J].
DAS, IJ ;
KAHN, FM .
MEDICAL PHYSICS, 1989, 16 (03) :367-375