Auditory cortical images of cochlear-implant stimuli: Dependence on electrode configuration

被引:134
作者
Bierer, JA
Middlebrooks, JC
机构
[1] Univ Michigan, Kresge Hearing Res Inst, Dept Otorhinolaryngol, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Program Neurosci, Ann Arbor, MI 48109 USA
关键词
D O I
10.1152/jn.00212.2001
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
This study examines patterns of auditory cortical activity elicited by single-pulse cochlear implant stimuli that vary in electrode configuration, cochlear place of stimulation, and stimulus level. Recordings were made from the primary auditory cortex (area A1) of ketamine-anesthetized guinea pigs. The spatiotemporal pattern of neural spike activity was measured simultaneously across 16 cortical locations spanning approximately 2-3 octaves of the tonotopic axis. Such a pattern, averaged over 40 presentations of any particular stimulus, was defined as the "cortical image" of that stimulus. Acutely deafened guinea pigs were implanted with a 6-electrode animal version of the 22-electrode Nucleus banded electrode array (Cochlear). Cochlear electrode configurations consisted of monopolar (MP), bipolar (BP + N) with N inactive electrodes between the active and return electrodes (0 less than or equal to N less than or equal to 4), tripolar (TP) with one active electrode and two flanking return electrodes, and common ground (CG) with one active electrode and as many as five return electrodes. Cortical images typically showed a focus of maximum spike probability and minimum latency. Spike probabilities tended to decrease, and latencies tended to increase, with increasing cortical distance from that focus. Cortical images of TP stimuli were the most spatially compact, followed by BP + N images, and then MP images, which were the broadest. Images of CG stimuli were rather variable across animals and stimulus channels. The locations of cortical images shifted systematically from caudal to rostral as the cochlear place of stimulation changed from basal to apical. At the most sensitive cortical site for each condition, the dynamic ranges over which spike rates increased with increased current level were restricted to about 1-2 dB, regardless of configuration. Dynamic ranges tended to increase with increasing cortical distance from the most sensitive site. Electrode configurations that produced compact cortical images (e.g., TP and BP + 0) showed the greatest range of thresholds within each cortical image and the largest dynamic range at cortical sites removed from the most sensitive site.
引用
收藏
页码:478 / 492
页数:15
相关论文
共 46 条
[1]   Auditory cortical images of tones and noise bands [J].
Arenberg, JG ;
Furukawa, S ;
Middlebrooks, JC .
JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2000, 1 (02) :183-194
[2]   Field patterns in a 3D tapered spiral model of the electrically stimulated cochlea [J].
Briaire, JJ ;
Frijns, JHM .
HEARING RESEARCH, 2000, 148 (1-2) :18-30
[3]   PITCH PERCEPTION FOR DIFFERENT MODES OF STIMULATION USING THE COCHLEAR MULTIPLE-ELECTRODE PROSTHESIS [J].
BUSBY, PA ;
WHITFORD, LA ;
BLAMEY, PJ ;
RICHARDSON, LM ;
CLARK, GM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 95 (05) :2658-2669
[4]   Effects of stimulation mode on threshold and loudness growth in multielectrode cochlear implants [J].
Chatterjee, M .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 105 (02) :850-860
[5]   Optical imaging of cat auditory cortex cochleotopic selectivity evoked by acute electrical stimulation of a multi-channel cochlear implant [J].
Dinse, HR ;
Godde, B ;
Hilger, T ;
Reuter, G ;
Cords, SM ;
Lenarz, T ;
vonSeelen, W .
EUROPEAN JOURNAL OF NEUROSCIENCE, 1997, 9 (01) :113-119
[6]   PERFORMANCE OF PLANAR MULTISITE MICROPROBES IN RECORDING EXTRACELLULAR SINGLE-UNIT INTRACORTICAL ACTIVITY [J].
DRAKE, KL ;
WISE, KD ;
FARRAYE, J ;
ANDERSON, DJ ;
BEMENT, SL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (09) :719-732
[7]  
EDDINGTON D K, 1978, Annals of Otology Rhinology and Laryngology, V87, P5
[8]   Coding of sound-source location by ensembles of cortical neurons [J].
Furukawa, S ;
Xu, L ;
Middlebrooks, JC .
JOURNAL OF NEUROSCIENCE, 2000, 20 (03) :1216-1228
[9]  
GANTZ BJ, 1988, LARYNGOSCOPE, V98, P1100
[10]   A COCHLEAR FREQUENCY-POSITION FUNCTION FOR SEVERAL SPECIES - 29 YEARS LATER [J].
GREENWOOD, DD .