Validation of a new noninvasive, method (contrast-enhanced transthoracic second harmonic echo Doppler) for the evaluation of coronary flow reserve - Comparison with intracoronary Doppler flow wire

被引:182
作者
Caiati, C
Montaldo, C
Zedda, N
Montisci, R
Ruscazio, V
Lai, G
Cadeddu, J
Meloni, L
Iliceto, S
机构
[1] Univ Cagliari, Inst Cardiol, Cagliari, Italy
[2] S Maugeri Fdn, IRCCS Med Ctr Rehabil, Div Cardiol, Cassano Murge, Italy
关键词
D O I
10.1016/S0735-1097(99)00342-3
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
OBJECTIVES We tested the hypothesis that coronary flow reserve (CFR) in the left anterior descending coronary artery (LAD) as assessed by a new noninvasive method (contrast-enhanced transthoracic second harmonic echo Doppler) is in agreement with CFR measurements assessed by intracoronary Doppler flow wire. BACKGROUND Contrast-enhanced transthoracic second harmonic echo Doppler is a novel noninvasive method to detect blood flow velocity and reserve in the LAD. However, it has not yet been validated versus a gold-standard method. METHODS Twenty-five patients undergoing CFR assessment in the LAD by Doppler flow wire were also evaluated by contrast-enhanced transthoracic Doppler to record blood flow in the distal LAD at rest and during hyperemia obtained by adenosine IV infusion. In five patients CFR was evaluated twice (before and after angioplasty). RESULTS As a result of the combined use of IV contrast and second harmonic Doppler technology, feasibility in assessing coronary flow reserve equaled 100%. The agreement between the two methods was high. In fact, in all but five patients the maximum difference between the two CFR measurements was 0.38. Overall, the prediction (95%) interval of individual differences was -0.69 to +0.72. Reproducibility of CFR measurements was also high. The limits of the agreement (95%) between the two measurements were -0.32 to +0.32. CONCLUSIONS Coronary flow reserve in the LAD as assessed by contrast-enhanced transthoracic echo Doppler along with harmonic mode concurs very closely with Doppler flow wire CFR measurements. This new noninvasive method allows feasible, reliable and reproducible assessment of CFR in the LAD. (J Am Coll Cardiol 1999;34:1193-200) (C) 1999 by the American College of Cardiology.
引用
收藏
页码:1193 / 1200
页数:8
相关论文
共 21 条
[1]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]
Contrast-enhanced transthoracic second harmonic echo Doppler with adenosine - A noninvasive, rapid and effective method for coronary flow reserve assessment [J].
Caiati, C ;
Zedda, N ;
Montaldo, C ;
Montisci, R ;
Iliceto, S .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 34 (01) :122-130
[3]
Improved Doppler detection of proximal left anterior descending coronary artery stenosis after intravenous injection of a lung-crossing contrast agent: A transesophageal Doppler echocardiographic study [J].
Caiati, C ;
Aragona, P ;
Iliceto, S ;
Rizzon, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 27 (06) :1413-1421
[4]
New noninvasive method for coronary flow reserve assessment - Contrast-enhanced transthoracic second harmonic echo Doppler [J].
Caiati, C ;
Montaldo, C ;
Zedda, N ;
Bina, A ;
Iliceto, S .
CIRCULATION, 1999, 99 (06) :771-778
[5]
ASSESSMENT OF CORONARY-ARTERY DISEASE SEVERITY BY POSITRON EMISSION TOMOGRAPHY - COMPARISON WITH QUANTITATIVE ARTERIOGRAPHY IN 193 PATIENTS [J].
DEMER, LL ;
GOULD, KL ;
GOLDSTEIN, RA ;
KIRKEEIDE, RL ;
MULLANI, NA ;
SMALLING, RW ;
NISHIKAWA, A ;
MERHIGE, ME .
CIRCULATION, 1989, 79 (04) :825-835
[6]
Assessing the Hemodynamic Significance of Coronary Artery Stenoses: Analysis of Translesional Pressure-Flow Velocity Relations in Patients [J].
Donohue, Thomas J. ;
Kern, Morton J. ;
Aguirre, Frank V. ;
Bach, Richard G. ;
Wolford, Thomas ;
Bell, Calvin A. ;
Segal, Jerome .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (02) :449-458
[7]
VALIDATION OF A DOPPLER GUIDE WIRE FOR INTRAVASCULAR MEASUREMENT OF CORONARY-ARTERY FLOW VELOCITY [J].
DOUCETTE, JW ;
CORL, PD ;
PAYNE, HM ;
FLYNN, AE ;
GOTO, M ;
NASSI, M ;
SEGAL, J .
CIRCULATION, 1992, 85 (05) :1899-1911
[9]
ESTIMATION OF CORONARY FLOW RESERVE BY INTRACORONARY DOPPLER FLOW PROBES AND DIGITAL ANGIOGRAPHY [J].
GRAHAM, SP ;
COHEN, MD ;
HODGSON, JM .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 19 (03) :214-221
[10]
IMPROVED DOPPLER SIGNAL INTENSITY IN CORONARY-ARTERIES AFTER INTRAVENOUS PERIPHERAL INJECTION OF A LUNG-CROSSING CONTRAST AGENT (SHU-508A) [J].
ILICETO, S ;
CAIATI, C ;
ARAGONA, P ;
VERDE, R ;
SCHLIEF, R ;
RIZZON, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (01) :184-190