On-line removal of interferences via anion-exchange column separation for the determination of germanium, arsenic and selenium in biological samples by inductively coupled plasma mass spectrometry

被引:29
作者
Ko, FH [1 ]
Yang, MH [1 ]
机构
[1] NATL TSING HUA UNIV,DEPT NUCL SCI,HSINCHU 30043,TAIWAN
关键词
germanium; arsenic; selenium; inductively coupled plasma mass spectrometry; on-line removal of interference;
D O I
10.1039/ja9961100413
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method was developed for the determination by inductively coupled plasma mass spectrometry of germanium, arsenic and selenium in urine and blood after microwave acid digestion and on-line anion-exchange separation, After acid digestion, the analytes exist mainly as H2GeO3, H3AsO3, H3AsO4, H2SeO3 and H2SeO4 in the resulting sample solution. Separation of the analytes is achieved by adsorption of the analytes as anionic species on the column in relatively alkaline media with subsequent desorption of the analytes from the column as undissociated neutral species in acidic media, By loading the digested sample, adjusted to about pH 11.5, onto the column of the on-line separation system, the analytes as well as Cl and S, which may cause spectral interference in the inductively coupled plasma mass spectrometric determination, are retained on the column, In the subsequent elution step with eluent of about pH 1.6, the analytes are desorbed from the column while Cl and S are still retained. With this on-line analytical system, the determination of Ge-72 and As-75 could be completely free from the interference caused by (ClCl)-Cl-35-Cl-37 and (ArCl)-Ar-40-Cl-35, respectively. By following the established microwave digestion and on-line separation process, the detection limits obtained for Ge, As and Se were as low as 0.07, 0.13 and 0.3 ng ml(-1), respectively, The analytical reliability of the proposed method was evaluated by analysing commercially available urine and whole blood standards, and the results obtained were found to be in reasonably good agreement with the literature values.
引用
收藏
页码:413 / 420
页数:8
相关论文
共 39 条
[1]   REDUCTION OF POLYATOMIC ION INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY BY CRYOGENIC DESOLVATION [J].
ALVES, LC ;
WIEDERIN, DR ;
HOUK, RS .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1164-1169
[2]   RAPID ELECTRODIALYTIC CLEANUP OF BIOLOGICAL SAMPLES FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
BREWSTER, JD ;
PIOTROWSKI, EG .
JOURNAL OF CHROMATOGRAPHY, 1991, 585 (02) :213-217
[3]   DETERMINATION OF SELENIUM BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY UTILIZING A NEW HYDRIDE GENERATION SAMPLE INTRODUCTION SYSTEM [J].
BUCKLEY, WT ;
BUDAC, JJ ;
GODFREY, DV ;
KOENIG, KM .
ANALYTICAL CHEMISTRY, 1992, 64 (07) :724-729
[4]   DEVELOPMENT OF A PROCEDURE FOR THE DETERMINATION OF CHROMIUM IN SAMPLES OF URINE AND SERUM BY NEUTRON-ACTIVATION ANALYSIS [J].
BUTTNER, I ;
HAMM, V ;
KNOCHEL, A ;
SENGUPTA, R .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 346 (04) :446-452
[5]   COLUMN-SWITCHING TECHNIQUES FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF DRUGS IN BIOLOGICAL SAMPLES [J].
CAMPINSFALCO, P ;
HERRAEZHERNANDEZ, R ;
SEVILLANOCABEZA, A .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1993, 619 (02) :177-190
[6]  
CHEN CJ, 1995, THESIS NATIONAL TSIN
[7]   COMPARISON OF 2 DIGESTION METHODS FOR THE DETERMINATION OF SELENIUM IN BIOLOGICAL SAMPLES [J].
DUCROS, V ;
RUFFIEUX, D ;
BELIN, N ;
FAVIER, A .
ANALYST, 1994, 119 (08) :1715-1717
[8]   INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY FOR BIOLOGICAL ANALYSIS [J].
DURRANT, SF .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1992, 11 (02) :68-73
[9]   INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY - A REVIEW [J].
EVANS, EH ;
GIGLIO, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) :1-18
[10]   SIMPLE APPROACH TO REDUCING POLYATOMIC ION INTERFERENCES ON ARSENIC AND SELENIUM IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EVANS, EH ;
EBDON, L .