Cone beam CT guidance provides superior accuracy for complex needle paths compared with CT guidance

被引:31
作者
Busser, W. M. H. [1 ]
Braak, S. J. [2 ]
Futterer, J. J. [1 ]
van Strijen, M. J. L. [2 ]
Hoogeveen, Y. L. [1 ]
de Lange, F. [1 ]
Kool, L. J. Schultze [1 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Dept Radiol, NL-6525 ED Nijmegen, Netherlands
[2] St Antonius Hosp, Dept Radiol, Nieuwegein, Netherlands
关键词
SYSTEM DIAGNOSTIC-ACCURACY; RADIATION-EXPOSURE; GUIDED PROCEDURES; LUNG NODULES; ARM CT; BIOPSY; TIME; INTERVENTIONS; ABLATION; ANGIOGRAPHY;
D O I
10.1259/bjr.20130310
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Objective: To determine the accuracy of cone beam CT (CBCT) guidance and CT guidance in reaching small targets in relation to needle path complexity in a phantom. Methods: CBCT guidance combines three-dimensional CBCT imaging with fluoroscopy overlay and needle planning software to provide real-time needle guidance. The accuracy of needle positioning, quantified as deviation from a target, was assessed for inplane, angulated and double angulated needle paths. Four interventional radiologists reached four targets along the three paths using CBCT and CT guidance. Accuracies were compared between CBCT and CT for each needle path and between the three approaches within both modalities. The effect of user experience in CBCT guidance was also assessed. Results: Accuracies for CBCT were significantly better than CT for the double angulated needle path (2.2 vs 6.7 mm, p<0.001) for all radiologists. CBCT guidance showed no significant differences between the three approaches. For CT, deviations increased with increasing needle path complexity from 3.3 mm for the inplane placements to 4.4 mm (p=0.007) and 6.7 mm (p<0.001) for the angulated and double angulated CT-guided needle placements, respectively. For double angulated needle paths, experienced CBCT users showed consistently higher accuracies than trained users [1.8 mm (range 1.2-2.2) vs 3.3 mm (range 2.1-7.2) deviation from target, respectively; p=0.003]. Conclusion: In terms of accuracy, CBCT is the preferred modality, irrespective of the level of user experience, for more difficult guidance procedures requiring double angulated needle paths as in oncological interventions. Advances in knowledge: Accuracy of CBCT guidance has not been discussed before. CBCT guidance allows accurate needle placement irrespective of needle path complexity. For angulated and double-angulated needle paths, CBCT is more accurate than CT guidance.
引用
收藏
页数:6
相关论文
共 27 条
[1]
Multimodality Image Fusion-Guided Procedures: Technique, Accuracy, and Applications [J].
Abi-Jaoudeh, Nadine ;
Kruecker, Jochen ;
Kadoury, Samuel ;
Kobeiter, Hicham ;
Venkatesan, Aradhana M. ;
Levy, Elliot ;
Wood, Bradford J. .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 35 (05) :986-998
[2]
Principles of and Advances in Percutaneous Ablation [J].
Ahmed, Muneeb ;
Brace, Christopher L. ;
Lee, Fred T., Jr. ;
Goldberg, S. Nahum .
RADIOLOGY, 2011, 258 (02) :351-369
[3]
Real-Time 3D Fluoroscopy Guidance During Needle Interventions: Technique, Accuracy, and Feasibility [J].
Braak, S. J. ;
van Strijen, M. J. L. ;
van Leersum, M. ;
van Es, H. W. ;
van Heesewijk, J. P. M. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (05) :W445-W451
[4]
3D cone-beam CT guidance, a novel technique in renal biopsy-results in 41 patients with suspected renal masses [J].
Braak, Sicco J. ;
van Melick, Harm H. E. ;
Onaca, Mircea G. ;
van Heesewijk, Johannes P. M. ;
van Strijen, Marco J. L. .
EUROPEAN RADIOLOGY, 2012, 22 (11) :2547-2552
[5]
Effective Dose during Needle Interventions: Cone-beam CT Guidance Compared with Conventional CT Guidance [J].
Braak, Sicco J. ;
van Strijen, Marco J. L. ;
van Es, Hendrik W. ;
Nievelstein, Rutger A. J. ;
van Heesewijk, Johannes P. M. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 22 (04) :455-461
[6]
Percutaneous Radiofrequency Ablation of Osteoid Osteomas with Use of Real-Time Needle Guidance for Accurate Needle Placement: A Pilot Study [J].
Busser, Wendy M. H. ;
Hoogeveen, Yvonne L. ;
Veth, Rene P. H. ;
Schreuder, H. W. Bart ;
Balguid, Angelique ;
Renema, W. KlaasJan ;
SchultzeKool, Leo J. .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 34 (01) :180-183
[7]
CT AND SONOGRAPHICALLY GUIDED NEEDLE-BIOPSY - CURRENT TECHNIQUES AND NEW INNOVATIONS [J].
CHARBONEAU, JW ;
READING, CC ;
WELCH, TJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1990, 154 (01) :1-10
[8]
Transthoracic needle biopsy using a C-arm cone-beam CT system: diagnostic accuracy and safety [J].
Choi, M. J. ;
Kim, Y. ;
Hong, Y. S. ;
Shim, S. S. ;
Lim, S. M. ;
Lee, J. K. .
BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1014) :E182-E187
[9]
Percutaneous transthoracic needle biopsy of small (aparts per thousandcurrency sign1 cm) lung nodules under C-arm cone-beam CT virtual navigation guidance [J].
Choo, Ji Yung ;
Park, Chang Min ;
Lee, Nyoung Keun ;
Lee, Sang Min ;
Lee, Hyun-Ju ;
Goo, Jin Mo .
EUROPEAN RADIOLOGY, 2013, 23 (03) :712-719
[10]
Thermal ablation therapy for focal malignancy: A unified approach to underlying principles, techniques, and diagnostic imaging guidance [J].
Goldberg, SN ;
Gazelle, GS ;
Mueller, PR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (02) :323-331