Monte Carlo and analytical calculation of proton pencil beams for computerized treatment plan optimization

被引:62
作者
Carlsson, AK
Andreo, P
Brahme, A
机构
[1] UNIV STOCKHOLM,S-17176 STOCKHOLM,SWEDEN
[2] INT ATOM ENERGY AGCY,DOSIMETRY SECT,RIHU,A-1400 VIENNA,AUSTRIA
关键词
D O I
10.1088/0031-9155/42/6/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Proton pencil beams in water, in a format suitable for treatment planning algorithms and covering the radiotherapy energy range (50-250 MeV), have been calculated using a modified version of the Monte Carlo code PTRAN. A simple analytical model has also been developed for calculating proton broad-beam dose distributions which is in excellent agreement with the Monte Carlo calculations. Radial dose distributions are also calculated analytically and narrow proton pencil-beam dose distributions derived. The physical approximations in the Monte Carlo code and in the analytical model together with their limitations are discussed. Examples showing the use of the calculated set of proton pencil beams as input to an existing photon treatment planning algorithm based on biological optimization are given for fully 3D scanned proton pencil beams; these include intensity modulated beams with range shift and scanning in the transversal plane.
引用
收藏
页码:1033 / 1053
页数:21
相关论文
共 62 条
[1]   OPTIMIZATION OF UNCOMPLICATED CONTROL FOR HEAD AND NECK TUMORS [J].
AGREN, A ;
BRAHME, A ;
TURESSON, I .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 19 (04) :1077-1085
[2]  
AGRENCRONQVIST AK, 1995, THESIS U STOCKHOLM
[3]   A PENCIL BEAM MODEL FOR PHOTON DOSE CALCULATION [J].
AHNESJO, A ;
SAXNER, M ;
TREPP, A .
MEDICAL PHYSICS, 1992, 19 (02) :263-273
[4]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[5]   CALCULATION AND APPLICATION OF POINT SPREAD FUNCTIONS FOR TREATMENT PLANNING WITH HIGH-ENERGY PHOTON BEAMS [J].
AHNESJO, A ;
ANDREO, P ;
BRAHME, A .
ACTA ONCOLOGICA, 1987, 26 (01) :49-56
[6]  
AHNESJO A, 1991, WORLD C MED PHYS BIO, P316
[7]  
AHNESJO A, 1984, 8TH P INT C US COMP, P17
[8]   MONTE-CARLO TECHNIQUES IN MEDICAL RADIATION PHYSICS [J].
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :861-920
[9]  
ANDREO P, 1988, MONTE CARLO TRANSPOR, P437
[10]  
[Anonymous], DOSIMETRY IONIZING R