Arc-deposited hydroxypatite-coated cups

被引:10
作者
Capello, WN
D'Antonio, JA
Manley, MT
Feinberg, JR
机构
[1] Indiana Univ, Sch Med, Dept Orthopaed Surg, Indianapolis, IN 46202 USA
[2] Greater Pittsburgh Orthoped Associates, Moon Township, PA USA
[3] Stryker Orthopaed, Mahwah, NJ USA
关键词
D O I
10.1097/01.blo.0000191276.88151.84
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A second-generation arc-deposited hydroxyapatite-coated acetabular component was developed after detailed analysis of a failed first-generation grit-blasted hydroxyapatite-coated cup. In our prospective multicenter study we aimed to compare clinical results and survivorship of the newly designed cup to the previous cup and to a similarly designed porous-coated cup during the same time period. T e effect of bearing surface-alumina-on-alumina to polyethylene-on-metal - was analyzed. Consecutive patients were assigned randomly to one of three groups: arc-deposited hydroxyapatite-coated cup with alumina-on-alumina bearing surface; porous-coated cup with alumina-on-alumina bearing surface; or porous-coated cup with polyethylene-on-metal bearing surface. A fourth comparison group with a grit-blasted hydroxyapatite-coated cup and polyethylene-on-metal bearing surface was culled from a previous study. Acetabular mechanical failure rate is 0.0%, 0.0%, 1.9%, and 11.2% for the arc-deposited hydroxyapatite-coated, porous-coated with alumina bearing surface, porous-coated with polyethylene-on-metal bearing, and grit-blasted hydroxyapatite-coated cup, respectively, at 4- to 7-years' followup. There are no cases of acetabular osteolysis or femoral neck scalloping and no radiolucent lines in any zone with the arc-deposited hydroxyapatite-coated cup. The cups with the alumina-onalumina bearing surface had less scalloping of the femoral neck than those with a polyethylene-on-metal bearing surface. Early results with this second-generation arc-deposited hydroxyapatite-coated cup are promising. Level of Evidence: Therapeutic study, Level I (high-quality randomized trial with statistically significant difference or no statistically significant difference but narrow confidence intervals). See the Guidelines for Authors for a complete description of levels of evidence.
引用
收藏
页码:305 / 312
页数:8
相关论文
共 29 条
[1]
FAILURE OF THE MECRING SCREW-RING ACETABULAR COMPONENT IN TOTAL HIP-ARTHROPLASTY - A 3-YEAR TO 7-YEAR FOLLOW-UP-STUDY [J].
BRUIJN, JD ;
SEELEN, JL ;
FEENSTRA, RM ;
HANSEN, BE ;
BERNOSKI, FP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (05) :760-766
[2]
FAILURE OF THE MECRON SCREW-IN RING [J].
CAPELLO, WN ;
COLYER, RA ;
KERNEK, CB ;
CARNAHAN, JV ;
HESS, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (05) :835-836
[3]
The Harris-Galante porous-coated acetabular component with screw fixation - An average ten-year follow-up study [J].
Clohisy, JC ;
Harris, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (01) :66-73
[4]
Survival and polyethylene wear of porous-coated acetabular components in patients less than fifty years old - Results at nine to fourteen years [J].
Crowther, JD ;
Lachiewicz, PF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (05) :729-735
[5]
D'Antonio JA, 2001, CLIN ORTHOP RELAT R, P101
[6]
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[7]
POROUS-COATED HIP-REPLACEMENT - THE FACTORS GOVERNING BONE INGROWTH, STRESS SHIELDING, AND CLINICAL-RESULTS [J].
ENGH, CA ;
BOBYN, JD ;
GLASSMAN, AH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (01) :45-55
[8]
CEMENTLESS ACETABULAR COMPONENTS [J].
ENGH, CA ;
GRIFFIN, WL ;
MARX, CL .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (01) :53-59
[9]
A 10-year minimum follow-up of hydroxyapatite-coated threaded cups - Clinical, radiographic and Survivorship analyses with comparison to the literature [J].
Epinette, JA ;
Manley, MT ;
D'Antonio, JA ;
Edidin, AA ;
Capello, WN .
JOURNAL OF ARTHROPLASTY, 2003, 18 (02) :140-148
[10]
Cementless acetabular fixation at fifteen years - A comparison with the same surgeon's results following acetabular fixation with cement [J].
Gaffey, JL ;
Callaghan, JJ ;
Pedersen, DR ;
Goetz, DD ;
Sullivan, PM ;
Johnston, RC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (02) :257-261