Comparison of fat suppression strategies in 3D spiral coronary magnetic resonance angiography

被引:17
作者
Börnert, P [1 ]
Stuber, M [1 ]
Botnar, RM [1 ]
Kissinger, KV [1 ]
Manning, WJ [1 ]
机构
[1] Philips Res Labs, D-22335 Hamburg, Germany
关键词
coronary angiography; MR imaging; fat suppression; spiral imaging; magnetic resonance angiography;
D O I
10.1002/jmri.10078
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In the present study, the impact of the two different fat suppression techniques was investigated for free breathing 3D spiral coronary magnetic resonance angiography (MRA). As the coronary arteries are embedded in epicardial fat and are adjacent to myocardial tissue, magnetization preparation such as T-2-preparation and fat suppression is essential for coronary discrimination. Material and Methods: Fat-signal suppression in three-dimensional (3D) thin-slab coronary MRA based on a spiral k-space data acquisition can either be achieved by signal pre-saturation using a spectrally selective inversion recovery pre-pulse or by spectral-spatial excitation. In the present study, the performance of the two different approaches was studied in healthy subjects. Results: No significant objective or subjective difference was found between the two fat suppression approaches. Conclusion: Spectral pre-saturation seems preferred for coronary MRA applications due to the ease of implementation and the shorter cardiac acquisition window.
引用
收藏
页码:462 / 466
页数:5
相关论文
共 24 条
[1]  
Börnert P, 2001, MAGNET RESON MED, V45, P172, DOI 10.1002/1522-2594(200101)45:1<172::AID-MRM1024>3.0.CO
[2]  
2-L
[3]   Improved coronary artery definition with T2-weighted, free-breathing, three-dimensional coronary MRA [J].
Botnar, RM ;
Stuber, M ;
Danias, PG ;
Kissinger, KV ;
Manning, WJ .
CIRCULATION, 1999, 99 (24) :3139-3148
[4]   CORONARY ANGIOGRAPHY WITH MAGNETIZATION-PREPARED T-2 CONTRAST [J].
BRITTAIN, JH ;
HU, BS ;
WRIGHT, GA ;
MEYER, CH ;
MACOVSKI, A ;
NISHIMURA, DG .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) :689-696
[5]   H-1-NMR CHEMICAL-SHIFT SELECTIVE (CHESS) IMAGING [J].
HAASE, A ;
FRAHM, J ;
HANICKE, W ;
MATTHAEI, D .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (04) :341-344
[6]   FAST 3-DIMENSIONAL MAGNETIC-RESONANCE-IMAGING [J].
IRARRAZABAL, P ;
NISHIMURA, DG .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) :656-662
[7]   Impact of bulk cardiac motion on right coronary MR angiography and vessel wall Imaging [J].
Kim, WY ;
Stuber, M ;
Kissinger, KV ;
Andersen, NT ;
Manning, WJ ;
Botnar, RM .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (04) :383-390
[8]   CORONARY-ARTERIES - 3-DIMENSIONAL MR-IMAGING WITH FAT SATURATION AND MAGNETIZATION TRANSFER CONTRAST [J].
LI, D ;
PASCHAL, CB ;
HAACKE, EM ;
ADLER, LP .
RADIOLOGY, 1993, 187 (02) :401-406
[9]   Multifrequency interpolation for fast off-resonance correction [J].
Man, LC ;
Pauly, JM ;
Macovski, A .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (05) :785-792
[10]   A PRELIMINARY-REPORT COMPARING MAGNETIC-RESONANCE CORONARY ANGIOGRAPHY WITH CONVENTIONAL ANGIOGRAPHY [J].
MANNING, WJ ;
LI, W ;
EDELMAN, RR .
NEW ENGLAND JOURNAL OF MEDICINE, 1993, 328 (12) :828-832