European Concerted Action on Anticoagulation (ECAA): multicentre international sensitivity index calibration of two types of point-of-care prothrombin time monitor systems

被引:29
作者
Poller, L [1 ]
Keown, M
Chauhan, N
van den Besselaar, AMHP
Tripodi, A
Jespersen, J
Shiach, C
Horellou, MH
Dias, D
Egberg, N
Iriarte, JA
Otridge, B
机构
[1] Univ Manchester, Sch Biol Sci, ECAA Cent Facil, Manchester, Lancs, England
[2] Leiden Univ, Med Ctr, Haemostasis & Thrombosis Res Ctr, Leiden, Netherlands
[3] Univ Milan, IRCCS, Maggiore Hosp, A Bianchi Bonomi Haemophilia & Thrombosis Ctr, Milan, Italy
[4] Univ So Denmark, Dept Thrombosis Res, Odense, Denmark
[5] Ribe Cty Hosp, Dept Clin Biochem, Esbjerg, Denmark
[6] Manchester Royal Infirm, Dept Haematol, Manchester M13 9WL, Lancs, England
[7] Hop Hotel Dieu, Cent Hematol Lab, Paris, France
[8] Hosp S Joao De Deus, Serv Imunohomoterapia, P-4760 Oporto, Portugal
[9] Karolinska Hosp, Dept Clin Chem, S-10401 Stockholm, Sweden
[10] Hosp Civil Basurto, Inst Epidemiol & Prevenc Enfermedades Cardiovasc, Bilbao, Spain
[11] Hippocrat Gen Hosp, Anticoagulant Unit, Athens, Greece
[12] Mater Misericordiae Univ Hosp, Dept Haematol, Dublin 7, Ireland
关键词
prothrombin time; international normalized ratio; international sensitivity index; multicentre calibration; point-of-care test monitors;
D O I
10.1046/j.0007-1048.2002.03361.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A multicentre modified World Health Organization (WHO)-type international sensitivity index (ISI) calibration has been performed at 10 European Concerted Action on Anticoagulation (ECAA) national laboratories using non-citrated whole-blood on two point-of-care test (POCT) prothrombin time (PT) monitor systems. CoaguChek Mini and TAS PT-NC, using single lots of test cards/strips. The relevant species (human and rabbit) WHO international reference preparations (IRPs) were tested with the manual PT technique on citrated plasma from the same blood donations. The ISI was calculated from the slope of the orthogonal regression line relating log PT (POCT) to log PT (IRP). The mean ISI of the CoaguChek Mini system was 1.75 and 1.13 with the prothrombin time non-citrated Thrombolytic Assessment System (TAS PT-NC). With the CoaguChek Mini system. seven out of 10 calibrations exceeded the current 3%, WHO recommended limit for the coefficient of variation (CV) of the slope with conventional PT testing, whereas with the TAS PT-NO system, it was eight out of 10. All the POCT calibrations had a CV of the slope <5%. It is suggested that this level of precision be adopted as the limit of acceptability of calibration of these monitor systems. In these circumstances. the modified WHO-type ISI calibration appeared to be satisfactory for the POCT whole-blood monitors.
引用
收藏
页码:844 / 850
页数:7
相关论文
共 14 条
[1]  
[Anonymous], 1984, THROMBOPLASTIN CALIB
[2]   Managing oral anticoagulant therapy [J].
Ansell, J ;
Hirsh, J ;
Dalen, J ;
Bussey, H ;
Anderson, D ;
Poller, L ;
Jacobson, A ;
Deykin, D ;
Matchar, D .
CHEST, 2001, 119 (01) :22S-38S
[3]  
HERMANS J, 1983, THROMB HAEMOSTASIS, V50, P712
[4]  
Poller L, 1996, THROMB HAEMOSTASIS, V76, P977
[5]  
THOMSON JM, 1986, THROMB HAEMOSTASIS, V55, P379
[6]  
THOMSON JM, 1984, THROMB HAEMOSTASIS, V52, P336
[7]  
TRIPODI A, 1994, THROMB HAEMOSTASIS, V72, P261
[8]  
Tripodi A, 1998, THROMB HAEMOSTASIS, V79, P439
[9]  
Tripodi A, 1997, THROMB HAEMOSTASIS, V78, P855
[10]  
TRIPODI A, 1993, THROMB HAEMOSTASIS, V70, P921