Determination of trimetazidine HCl by adsorptive stripping square-wave voltammetry at a glassy carbon electrode

被引:35
作者
Ghoneim, MM [1 ]
Khashaba, PY
Beltagi, AM
机构
[1] Tanta Univ, Fac Sci, Dept Chem, Tanta 31527, Egypt
[2] Assiut Univ, Fac Pharm, Dept Pharmaceut Analyt Chem, Assiut, Egypt
[3] Fac Educ, Kafr Al Sheikh 33516, Egypt
[4] Tanta Univ, Kafr Al Sheikh, Egypt
关键词
trimetazidine hydrochloride; square-wave adsorptive voltammetry; glassy carbon electrode; tablet analysis; human urine determination;
D O I
10.1016/S0731-7085(01)00525-8
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学]; 081704 [应用化学];
摘要
The adsorptive and electrochemical behavior of trimetazidine hydrochloride on a glassy carbon electrode were investigated in acetate buffer solution by using cyclic and square-wave voltammetry. Cyclic voltammetric studies indicated the oxidation of trimetazidine hydrochloride at the electrode surface through a single two-electron irreversible step and fundamentally controlled by adsorption. The solution condition and instrumental parameters were optimized for the determination of the authentic drug using adsorptive square wave stripping voltammetry. Trimetazidine hydrochloride gave a sensitive adsorptive oxidative peak at 0.750 V (vs. Ag/AgCl). The oxidation peak was used to determine authentic trimetazidine hydrochloride concentration in the range 5.0 x 10(-8)-5.0 x 10(-6) M with a detection limit of 2.0 x 10(-8) M. The procedure was successfully applied for assay of trimetazidine hydrochloride in the tablet dosage form (Vastarel). A mean recovery of 94.7% with a relative standard deviation (R.S.D.) of 0.88% was obtained. Applicability to assay the drug in urine samples was illustrated. The peak current was linear with the drug concentration in the range 17-85 mug per ml urine. The detection limit was 1.7 ug ml(-1) urine. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:235 / 241
页数:7
相关论文
共 11 条
[1]
Structure-property relationships of trimetazidine derivatives and model compounds as potential antioxidants [J].
Ancerewicz, J ;
Migliavacca, E ;
Carrupt, PA ;
Testa, B ;
Brée, F ;
Zini, R ;
Tillement, JP ;
Labidalle, S ;
Guyot, D ;
Chauvet-Monges, AM ;
Crevat, A ;
Le Ridant, A .
FREE RADICAL BIOLOGY AND MEDICINE, 1998, 25 (01) :113-120
[2]
Bari V. R., 1999, INDIAN DRUGS, V36, P289
[3]
TRACE DETERMINATION OF TRIMETAZIDINE IN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING FLUORESCENCE DETECTION [J].
COURTE, S ;
BROMET, N .
JOURNAL OF CHROMATOGRAPHY, 1981, 224 (01) :162-167
[4]
de Leiris J, 1990, Rev Port Cardiol, V9, P153
[5]
DETERMINATION OF TRIMETAZIDINE IN BIOLOGICAL-FLUIDS BY GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
FAY, L ;
MICHEL, G ;
GOUPIT, P ;
HARPEY, C ;
PROST, M .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1989, 490 (01) :198-205
[6]
Huskisson EC, 1996, BRIT J RHEUMATOL, V35, P29
[7]
JAKSON PJ, 1996, XENOBIOTICA, V26, P221
[8]
LAVIRON E, 1980, J ELECTROANAL CHEM, V112, P1, DOI 10.1016/S0022-0728(80)80002-7
[9]
PARFITT K, 1999, MARTINDALE COMPLETE, P959
[10]
Stability indicating HPTLC determination of trimetazidine as bulk drug and in pharmaceutical formulations [J].
Thoppil, SO ;
Cardoza, RM ;
Amin, PD .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2001, 25 (01) :15-20