The impact of cardiology on the collective effective dose in the North of England

被引:79
作者
Broadhead, DA [1 ]
Chapple, CL [1 ]
Faulkner, K [1 ]
Davies, ML [1 ]
McCallum, H [1 ]
机构
[1] S CLEVELAND HOSP,DEPT REG MED PHYS,MIDDLESBROUGH TS4 3BW,CLEVELAND,ENGLAND
关键词
D O I
10.1259/bjr.70.833.9227231
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Two cardiology X-ray rooms were monitored with dose-area product meters as part of a Regional Patient Dosimetry Programme. Dose-area product measurements on over 2000 patients undergoing examinations in the cardiology rooms are presented. The data have been corrected according to patient size where possible. In room A mean dose-area product values for coronary angiography, coronary angioplasty, radiofrequency ablation and mitral valvuloplasty were found to be 47.7, 72.2, 91.1 and 161.9 Gy cm(2) respectively. In room B mean dose-area product values for coronary angiography and coronary angioplasty were found to be 23.4 and 51.6 Gy cm(2) respectively. Observational studies were used to deduce the typical projections and technique factors. This typical examination was used to simulate an angiogram from which it was possible to derive factors to convert measured dose-area product values into estimates of effective dose. In room A: the effective doses were estimated to be 9.4, 14.2, 17.3 and 29.3 mSv for coronary angiography, coronary angioplasty, radiofrequency ablation and mitral valvuloplasty, respectively. The effective doses during coronary angiography and coronary angioplasty, performed in room B, were found to be 4.6 and 10.2 mSv, respectively. A regional survey of the frequency of these cardiac procedures was performed. It was deduced that the annual collective effective dose from these cardiac procedures in the North of England, the former Northern Region, was 45.7 manSv.
引用
收藏
页码:492 / 497
页数:6
相关论文
共 16 条
[1]   RADIATION DOSE TO PATIENT IN CARDIAC RADIOLOGY [J].
ARDRAN, GM ;
HAMILL, J ;
EMRYSROB.E .
BRITISH JOURNAL OF RADIOLOGY, 1970, 43 (510) :391-&
[2]   THE IMPACT OF DIGITAL IMAGING ON PATIENT DOSES DURING BARIUM STUDIES [J].
BROADHEAD, DA ;
CHAPPLE, CL ;
FAULKNER, K .
BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (813) :992-996
[3]   A PHANTOM BASED METHOD FOR DERIVING TYPICAL PATIENT DOSES FROM MEASUREMENTS OF DOSE-AREA PRODUCT ON POPULATIONS OF PATIENTS [J].
CHAPPLE, CL ;
BROADHEAD, DA ;
FAULKNER, K .
BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (814) :1083-1086
[4]   CORONARY ANGIOGRAPHY - AN ANALYSIS OF RADIOGRAPHIC PRACTICE IN THE UK [J].
COULDEN, RA ;
READMAN, LP .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (784) :327-331
[5]  
*DOS WORK PART I S, 1992, NAT PROT PAT DOS MEA
[6]   RADIATION-DOSES AND SOMATIC RISK TO PATIENTS DURING CARDIAC RADIOLOGICAL PROCEDURES [J].
FAULKNER, K ;
LOVE, HG ;
SWEENEY, JK ;
BARDSLEY, RA .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (700) :359-363
[7]  
FAULKNER K, 1992, RADIAT PROT DOSIM, V43, P131
[8]   A SURVEY OF RADIATION-DOSES TO PATIENTS IN 5 COMMON DIAGNOSTIC EXAMINATIONS .1. PRINCIPLES AND TECHNIQUES OF DOSE EVALUATION BY INDIRECT MEASUREMENTS .2. SURVEY RESULTS [J].
HARRISON, RM ;
CLAYTON, CB ;
DAY, MJ ;
OWEN, JP ;
YORK, MF .
BRITISH JOURNAL OF RADIOLOGY, 1983, 56 (666) :383-395
[9]  
HART D, 1994, NRPBR262
[10]   EFFECT OF PULSED PROGRESSIVE FLUOROSCOPY ON REDUCTION OF RADIATION-DOSE IN THE CARDIAC-CATHETERIZATION LABORATORY [J].
HOLMES, DR ;
WONDROW, MA ;
GRAY, JE ;
VETTER, RJ ;
FELLOWS, JL ;
JULSRUD, PR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (01) :159-162