Results of the NIST national ball plate round robin

被引:2
作者
Caskey, GW
Phillips, SD
Borchardt, BR
机构
[1] Natl. Inst. of Std. and Technology, Gaithersburg
[2] Precision Engineering Division, NIST Mfg. Engineering Laboratory, Natl. Inst. of Std. and Technology
关键词
ball plate; CMM; coordinate measuring machine; coordinate metrology; round robin;
D O I
10.6028/jres.102.008
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This report examines the results of the ball plate round robin administered by NIST. The round robin was part of an effort to assess the current state of industry practices for measurements made using coordinate measuring machines. Measurements of a two-dimensional ball plate (240 mm by 240 mm) on 41 coordinate measuring machines were collected and analyzed. Typically, the deviations of the reported X and Y coordinates from the calibrated values were within +/- 5 mu m, with some coordinate deviations exceeding 20.0 mu m. One of the most significant observations from these data was that over 75 % of the participants failed to correctly estimate their measurement error on one or more of the ball plate spheres.
引用
收藏
页码:85 / 93
页数:9
相关论文
共 6 条
[1]  
HOCKEN RJ, 1979, 791752 NBSIR NBS US
[2]  
*ISO, 1993, GUID EXP UNC MEAS
[3]  
KUNZMANN H, 1990, ANN CIRP, V39, P461
[4]  
PHILLIPS SD, 1994, CAL LAB, V1, P7
[5]  
REEVE CP, 1974, 74532 NBSIR NBS US
[6]  
Taylor B. N., 1994, 1297 NIST