Comparison of TEST 1 with SRS 100 and ICSH reference method for the measurement of the length of sedimentation reaction in blood

被引:16
作者
Ozdem, S [1 ]
Akbas, HS
Donmez, L
Gultekin, M
机构
[1] Akdeniz Univ, Fac Med, Cent Lab, Clin Biochem Unit, TR-07070 Antalya, Turkey
[2] Akdeniz Univ, Fac Med, Cent Lab, Microbiol Unit, TR-07070 Antalya, Turkey
[3] Akdeniz Univ, Fac Med, Cent Lab, Dept Publ Hlth, TR-07070 Antalya, Turkey
关键词
fibrinogen; hematocrit; length of sedimentation reaction; TEST; 1; Westergren method;
D O I
10.1515/CCLM.2006.075
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 [基础医学];
摘要
Background: We evaluated the measurement of length of sedimentation reaction in blood ( LSRB) by TEST 1 and compared the results with those for the Westergren and Sed Rate Screener 100 ( SRS 100) methods. Methods: LSRB was measured in 113 paired blood samples. Results: TEST 1 correlated significantly with the Westergren ( r=0.94) and SRS 100 ( r=0.90) methods with low bias ( -0.29 and -1.92 mm/h, respectively) and limits of agreement ( -14.5 to 13.9, and -23.4 to 19.6 mm/h, respectively). Hematocrit ( Htc) correlated negatively with LSRB in TEST 1 ( r=-0.54) and SRS 100 ( r=-0.53) only in samples with high Htc ( >= 35%). The bias and limits of agreement between TEST 1 and Westergren in samples with low ( -1.46 and -22.3 to 19.3 mm/h) and high ( 0.43 and -7.29 to 8.14 mm/h) Htc were comparable to those between SRS 100 and Westergren ( 1.83 and -27.2 to 30.9 mm/h for low, 0.71 and -7.27 to 8.70 mm/h for high Htc samples). Total protein and fibrinogen correlated similarly with LSRB in both TEST 1 ( r=0.23 and 0.48, respectively) and SRS 100 ( r=0.30 and 0.51, respectively). Conclusions: The findings suggested that TEST 1 is a reliable, precise and accurate system for measurement of LSRB in clinical laboratories with high workload.
引用
收藏
页码:407 / 412
页数:6
相关论文
共 21 条
[1]
[Anonymous], KLIIN LAB
[2]
[Anonymous], 1921, Acta. Med. Scand, DOI [10.1152/physrev.1929.9.2.241, DOI 10.1111/J.0954-6820.1921.TB15179.X]
[3]
Arezzini C, 1999, AM J CLIN PATHOL, V112, P722
[4]
BALLOU SP, 1996, TXB RHEUMATOLOGY, P699
[5]
ERYTHROCYTE SEDIMENTATION-RATE - FROM FOLKLORE TO FACTS [J].
BEDELL, SE ;
BUSH, BT .
AMERICAN JOURNAL OF MEDICINE, 1985, 78 (06) :1001-1009
[6]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]
BULL BS, 1993, J CLIN PATHOL, V46, P198
[8]
de Jonge N, 2000, CLIN CHEM, V46, P881
[9]
FABRY TL, 1987, BLOOD, V70, P1572
[10]
Giavarina D, 1999, AM J CLIN PATHOL, V112, P721