Central and juxta-endplate vertebral body screw placement - A biomechanical analysis in a human cadaveric model

被引:18
作者
Lowe, T
O'Brien, M
Smith, D
Fitzgerald, D
Vraney, R
Eule, J
Alongi, P
机构
[1] Woodridge Orthopaed & Spine Ctr, PC, Wheat Ridge, CO USA
[2] Orthopaed Spine Care Ctr LI, PC, Melville, NY USA
[3] M&M Orthopaed, Naperville, IL USA
关键词
D O I
10.1097/00007632-200202150-00009
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. In vitro biomechanical testing of transvertebral body screws in different positions in both axial pull-out and toggle. Objectives, To determine the relative strength of unicortical versus bicortical screw fixation within the vertebral body and to determine comparative strength of juxta-endplate and central screw positions with and without staples in both axial pull-out and toggle modes. Summary of Background Data. Loss of fixation is common in centrally placed screws at the rostral end of a construct. To preserve segmental vessels, juxta-endplate screw positions are often used. The biomechanical strength of such screw placement methods has not been measured. Methods. Eighty-three human cadaveric vertebral dies were tested for axial pull-out and toggle with and without staples. Screw positions included central, juxta and inferior endplate. Juxta-endplate screws were toggled in both the rostral and caudal directions perpendicular to the screw axes. Results. Unicortical fixation resulted in a 93% decrease In axial pull-out strength compared with bicortical fixation. Centrally placed screws and juxta-endplate screws were equivalent in axial pull-out if no staples were used. The juxta-endplate screw with a staple that was toggled away from the endplate had the highest yield strength, followed by the central screw with a staple, and then the juxta-end plate screw without a staple toggled away from the endplate. Conclusions. Bicortical fixation is much stronger than unicortical fixation. Centrally placed screws are significantly, stronger when used with a staple. When preservation of segmental vessels is desirable, juxta-endplate screws should be placed in such a manner that compressive forces are directed away from the endplate.
引用
收藏
页码:369 / 373
页数:5
相关论文
共 9 条
[1]
Ansell R H, 1968, J Biomech, V1, P279, DOI 10.1016/0021-9290(68)90023-7
[2]
BETZ RR, 2000, SPINE, V14, P115
[3]
CORRELATIONS BETWEEN SCREW HOLE PREPARATION, TORQUE OF INSERTION, AND PULLOUT STRENGTH FOR SPINAL SCREWS [J].
DAFTARI, TK ;
HORTON, WC ;
HUTTON, WC .
JOURNAL OF SPINAL DISORDERS, 1994, 7 (02) :139-145
[4]
GEORGE DC, 1991, SPINE, V16, P181, DOI 10.1097/00007632-199116020-00015
[5]
Strength of fixation of anterior vertebral body screws [J].
Horton, WC ;
Blackstock, SF ;
Norman, JT ;
Hill, CS ;
Feiertag, MA ;
Hutton, WC .
SPINE, 1996, 21 (04) :439-444
[6]
KORANYI E, 1970, CLIN ORTHOP RELAT R, P283
[7]
LOWE TG, 2001, PEDIAT SPINE SURG, P515
[8]
RYAN MD, 1986, CLIN ORTHOP RELAT R, P196
[9]
SCHATZKER J, 1975, CLIN ORTHOP RELAT R, P115