Whole-body three-dimensional optoacoustic tomography system for small animals

被引:252
作者
Brecht, Hans-Peter [1 ]
Su, Richard [1 ]
Fronheiser, Matthew [2 ]
Ermilov, Sergey A. [1 ]
Conjusteau, Andre [1 ]
Oraevsky, Alexander A. [1 ]
机构
[1] Fairway Med Technol Inc, Houston, TX 77099 USA
[2] Seno Med Instruments Inc, San Antonio, TX 78230 USA
关键词
small animal imaging; optoacoustic; image reconstruction; functional imaging; blood; vasculature; IN-VIVO; PHOTOACOUSTIC TOMOGRAPHY; CONTRAST AGENT; CANCER; BRAIN; BLOOD;
D O I
10.1117/1.3259361
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We develop a system for three-dimensional whole-body optoacoustic tomography of small animals for applications in preclinical research. The tomographic images are obtained while the objects of study (phantoms or mice) are rotated within a sphere outlined by a concave arc-shaped array of 64 piezocomposite transducers. Two pulsed lasers operating in the near-IR spectral range (755 and 1064 nm) with an average pulsed energy of about 100 mJ, a repetition rate of 10 Hz, and a pulse duration of 15 to 75 ns are used as optical illumination sources. During the scan, the mouse is illuminated orthogonally to the array with two wide beams of light from a bifurcated fiber bundle. The system is capable of generating images of individual organs and blood vessels through the entire body of a mouse with spatial resolution of similar to 0.5 mm. (C) 2009 Society of Photo-Optical Instrumentation Engineers. [DOI: 10.1117/1.3259361]
引用
收藏
页数:8
相关论文
共 35 条
[1]   Opto-acoustic tomography of breast cancer with arc-array-transducer [J].
Andreev, VG ;
Karabutov, AA ;
Solomatin, SV ;
Savateeva, EV ;
Aleynikov, V ;
Zhulina, YV ;
Fleming, RD ;
Oraevsky, AA .
BIOMEDICAL OPTOACOUSTICS, 2000, 3916 :36-47
[2]  
[Anonymous], 2003, BIOMEDICAL PHOTONICS
[3]   In vivo small-animal imaging using micro-CT and digital subtraction angiography [J].
Badea, C. T. ;
Drangova, M. ;
Holdsworth, D. W. ;
Johnson, G. A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (19) :R319-R350
[4]  
BRECHT H, 2009, P SPIE, V7177
[5]  
Brecht HPF, 2004, P ANN INT IEEE EMBS, V26, P375
[6]   High sensitivity of in vivo detection of gold nanorods using a laser optoacoustic imaging system [J].
Eghtedari, Mohammad ;
Oraevsky, Alexander ;
Copland, John A. ;
Kotov, Nicholas A. ;
Conjusteau, Andre ;
Motamedi, Massoud .
NANO LETTERS, 2007, 7 (07) :1914-1918
[7]  
ERMILOV SA, 2007, P SPIE, V6437
[8]   Laser optoacoustic imaging system for detection of breast cancer [J].
Ermilov, Sergey A. ;
Khamapirad, Tuenchit ;
Conjusteau, Andre ;
Leonard, Morton H. ;
Lacewell, Ron ;
Mehta, Ketan ;
Miller, Tom ;
Oraevsky, Alexander A. .
JOURNAL OF BIOMEDICAL OPTICS, 2009, 14 (02)
[9]   Optical tomographic imaging of small animals [J].
Hielscher, AH .
CURRENT OPINION IN BIOTECHNOLOGY, 2005, 16 (01) :79-88
[10]  
Iwaki T., 2001, COLOR ATLAS SECTIONA