Effect of temperature on ultrasonic properties of the calcaneus in situ

被引:34
作者
Nicholson, PHF
Bouxsein, ML
机构
[1] Beth Israel Deaconess Med Ctr, Orthopaed Biomech Lab, Dept Orthoped Surg, Boston, MA 02215 USA
[2] Harvard Univ, Sch Med, Boston, MA USA
关键词
calcaneus; quantitative ultrasound (QUS); temperature effects;
D O I
10.1007/s001980200122
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
To assess the dependence of calcaneal quantitative ultrasound (QUS) on foot temperature, a series of acoustic measurements were made in five cadaver feet in situ (all soft tissues retained) over a temperature range of 25 degreesC to 40 degreesC in steps of 5 degreesC. An implanted probe was used to measured temperature directly in the calcaneus itself. Ultrasound velocity decreased linearly with increasing temperature, with a mean thermal coefficient of -2.2 m/s/degreesC. In contrast, broadband ultrasonic attenuation (BUA) increased with temperature with a mean thermal coefficient of +0.75 dB/MHz/degreesC. We argue that the temperature trends in velocity are likely to be due to the influence of fat, present in the bone marrow and in the soft tissues, which has a negative thermal coefficient for acoustic velocity. The attenuation trends may arise, in part, from greater scattering losses inside the cancellous bone due to an increased acoustic impedance mismatch between trabeculae and fatty marrow at higher temperatures. These considerations suggest that the greatest temperature effects may be expected in patients with a high proportion of fat within the measured volume and/or low calcaneal bone density. Given the magnitude of the thermal coefficients observed, the clinical impact of temperature-related QUS errors is likely to be modest for diagnostic purposes but of greater significance in follow up studies.
引用
收藏
页码:888 / 892
页数:5
相关论文
共 19 条
[1]
BARKMANN R, 1998, J CLIN DENSITOM, V1, P94
[2]
ULTRASONIC ANALYSIS OF THE YOUNG MODULUS OF CORTICAL BONE [J].
BONFIELD, W ;
TULLY, AE .
JOURNAL OF BIOMEDICAL ENGINEERING, 1982, 4 (01) :23-27
[3]
Ultrasound measurement on the calcaneus: Influence of immersion time and rotation of the foot [J].
Chappard, C ;
Berger, G ;
Roux, C ;
Laugier, P .
OSTEOPOROSIS INTERNATIONAL, 1999, 9 (04) :318-326
[4]
EVANS JA, 1992, CURRENT RES OSTEOPOR, V2, P43
[6]
Haney WD, 1986, TISSUE CHARACTERIZAT, P15
[8]
KISHIMOTO T, 1958, ACUSTICA, V8, P179
[9]
INFLUENCE OF FAT ON ULTRASOUND MEASUREMENTS OF THE OS CALCIS [J].
KOTZKI, PO ;
BUYCK, D ;
HANS, D ;
THOMAS, E ;
BONNEL, F ;
FAVIER, F ;
MEUNIER, PJ ;
ROSSI, M .
CALCIFIED TISSUE INTERNATIONAL, 1994, 54 (02) :91-95
[10]
MORRIS R, 1998, J CLIN DENSITOM, V1, P104