Frictional heating of bearing materials tested in a hip joint wear simulator

被引:118
作者
Lu, Z
McKellop, H
机构
[1] J Vernon Luck Orthopaedic Research Center, Los Angeles Orthopaedic Hospital
[2] Departments of Orthopaedic and Biomédical Engineering, University of Southern California
关键词
wear; frictional heat; hip prosthesis; polyethylene; finite element; UHMW POLYETHYLENE; ACETABULAR CUPS; METAL;
D O I
10.1243/0954411971534728
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In a hip simulator wear test using bovine serum as a lubricant, the heat generated by ball-cup friction may cause precipitation of the proteins from the lubricant. The resultant accumulation of a solid layer of precipitated protein between the ball and cup could artificially protect the bearing surfaces from wear, in a manner that does not occur in vivo. Alternatively, the gradual depletion of the soluble proteins could interfere with their ability to act as boundary lubricants on the bearing surfaces, thereby artificially increasing the wear rate. Because the rate of protein precipitation may depend on the maximum temperature at the bearing surfaces during sliding, rather than the mean temperature of the bulk lubricant, this study determined the transient surface temperatures using an array of thermocouples embedded in acetabular cups of GUR 415 ultra-high molecular weight polyethylene (UHMWPE) and femoral balls of metal or ceramic, in conjunction with a finite element model of the temperature distribution. The prostheses were tested at one cycle/s under a Paul-type, physiological load profile with 2030 N maximum force, with the load cycle synchronized to the motion cycle. The steady state temperatures of the bulk lubricant were 38 degrees C for the zirconia balls, 36 degrees C for the cobalt-chromium and 33 degrees C for the alumina. However, the corresponding surface temperatures of the polyethylene, calculated with the finite element model, were 99 degrees C with zirconia ceramic, 60 degrees C with cobalt-chromium alloy, and 45 degrees C with alumina ceramic. The rank order of the surface temperatures corresponded to the relative amounts of protein that were precipitated in the test chambers during wear tests with these materials.
引用
收藏
页码:101 / 108
页数:8
相关论文
共 34 条
[1]  
AMSTUTZ HC, 1968, J BIOMED MATER RES, V3, P547
[2]  
[Anonymous], CLIN ORTHOP RELAT R, DOI DOI 10.1097/00003086-199608001-00013
[3]  
Benedict RobertP., 1977, FUNDAMENTALS TEMPERA, V2nd
[4]  
BRAGDON CR, 1996, T 5 WORLD BIOM C TOR, pI582
[5]  
Buly R L, 1992, J Arthroplasty, V7, P315, DOI 10.1016/0883-5403(92)90056-V
[6]  
CLARK IC, 1996, J APPL BIOMATER, V33, P275
[7]  
Cooper J. R., 1993, Clinical Materials, V14, P295, DOI 10.1016/0267-6605(93)90016-Z
[8]   BIREFRINGENT STUDIES OF POLYETHYLENE WEAR SPECIMENS AND ACETABULAR CUPS [J].
COOPER, JR ;
DOWSON, D ;
FISHER, J .
WEAR, 1991, 151 (02) :391-402
[9]   HEAT-TRANSFER ANALYSIS OF FRICTIONAL HEAT DISSIPATION DURING ARTICULATION OF FEMORAL IMPLANTS [J].
DAVIDSON, JA ;
GIR, S ;
PAUL, JP .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH-APPLIED BIOMATERIALS, 1988, 22 (A3) :281-309
[10]  
DAVIDSON JA, 1988, J BIOMED MATER RES-A, V22, P69