Absolute measurement of oxygen edge structure in the quantum efficiency of X-ray CCDs

被引:10
作者
Prigozhin, G [1 ]
Bautz, M [1 ]
Kissel, S [1 ]
Ricker, G [1 ]
Kraft, S [1 ]
Scholze, F [1 ]
Thornagel, R [1 ]
Ulm, G [1 ]
机构
[1] PHYS TECH BUNDESANSTALT,D-10587 BERLIN,GERMANY
关键词
D O I
10.1109/23.603787
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
As part of the program to calibrate X-ray CCD detectors for the AXAF mission we have performed high-spectral resolution, absolute measurements of CCD quantum efficiency using the synchrotron radiation and grating monochromator facilities of the PTB radiometry laboratory at BESSY. We present here detailed measurements of the absolute quantum efficiency in the spectral region surrounding the oxygen K absorption edge for both front- and back-illuminated MIT Lincoln Laboratory CCDs. The absolute scale of the measurements is established by reference to an Electrical Substitution Radiometer. Near-edge absorption structure produces variations as large as 40% in the detection efficiency of the front-illuminated detectors. We find structure at energies immediately below the K absorption edge which we attribute to resonant absorption. The amplitude of edge structures in the back-illuminated devices is, as expected, much smaller Finally, we demonstrate that edge structure can be reliably detected in CCD response using an electron impact source and a grating monochromator.
引用
收藏
页码:970 / 974
页数:5
相关论文
共 9 条
[1]   K-SHELL PHOTOABSORPTION COEFFICIENTS OF O2, CO2, CO, AND N2O [J].
BARRUS, DM ;
BLAKE, RL ;
BUREK, AJ ;
CHAMBERS, KC ;
PREGENZER, AL .
PHYSICAL REVIEW A, 1979, 20 (03) :1045-1061
[2]  
BURKE BE, 1993, P SOC PHOTO-OPT INS, V2006, P272, DOI 10.1117/12.162840
[3]  
HENKE BL, 1993, ATOMIC DATA NUCLEAR, V54
[4]   ONE GIGARAD PASSIVATING NITRIDED OXIDES FOR 100-PERCENT INTERNAL QUANTUM EFFICIENCY SILICON PHOTODIODES [J].
KORDE, R ;
CABLE, JS ;
CANFIELD, LR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (06) :1655-1659
[5]  
LEDERER T, 1995, P SOC PHOTO-OPT INS, V2519, P92, DOI 10.1117/12.211893
[6]  
Owens A, 1996, X-RAY SPECTROM, V25, P33, DOI 10.1002/(SICI)1097-4539(199601)25:1<33::AID-XRS135>3.3.CO
[7]  
2-3
[8]   Calibration of X-ray CCDs with an erect-field grating spectrometer in the 0.2-1.5 keV band. [J].
Prigozhin, G ;
Bautz, M ;
Gendreau, K ;
Ricker, G .
EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY VII, 1996, 2808 :260-270
[9]   PLANE GRATING MONOCHROMATOR BEAMLINE FOR VUV RADIOMETRY [J].
SCHOLZE, F ;
KRUMREY, M ;
MULLER, P ;
FUCHS, D .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1994, 65 (10) :3229-3232