Evaluation of the analytic performances of the new HPLC system HLC-723 G7 for the measurement of hemoglobin A1C

被引:20
作者
Terreni, A
Paleari, R
Caldini, A
Ognibene, A
Mosca, A
Messeri, G
机构
[1] Careggi Hosp, Cent Clin Biochem Lab, I-50139 Florence, Italy
[2] Univ Milan, LITA, Dept Biomed Sci & Technol, I-20090 Milan, Italy
关键词
glycohemoglobin; HbA(1C); HPLC; evaluation;
D O I
10.1016/S0009-9120(03)00116-4
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: The analytical performance of a new automated HPLC system, for the determination of HbA(1C) in blood (Tosoh HLC-723 G7), was studied. Design and methods: The study design included the evaluation of imprecision, linearity, interference and carryover. Comparison study was performed by comparing HbA(1C) results with those obtained from an established method (Bio-Rad Variant II). Results: Total imprecision was less than 1.34% and the results were linear up to 17.2% HbA(1C). The method showed a wide analytical range, and no carryover between specimens. Comparison study yielded, r = 0.989, Sy.x = 0.255, regression equation (y = 0.9895x-0.35); Bland-Altman plot showed a mean bias = - 0.43% HbA(1C) with confidence limits ranging from -0.48% to -0.38% HbA(1C). The presence of abnormal hemoglobin was clearly revealed, and no interference from labile HbA(1C) was apparent. Conclusion: The HLC-723 G7 instrument seems to be a reliable system for routine assay of HbA(1C). (C) 2003 The Canadian Society of Clinical Chemists. All rights reserved.
引用
收藏
页码:607 / 610
页数:4
相关论文
共 18 条
[1]  
*AM DIAB ASS, DIABETES CARE S1, V23, pS24
[2]  
Amer Diabet Assoc, 2000, DIABETES CARE, V23, pS32
[3]  
[Anonymous], 1986, EP7P NCCLS
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
BROUGHTON PMG, 1969, J CLIN PATHOL, V22, P278
[6]  
Bry L, 2001, CLIN CHEM, V47, P153
[7]   Influence of in vivo hemoglobin carbamylation on HbA1c measurements by various methods [J].
Chachou, A ;
Randoux, C ;
Millart, H ;
Chanard, J ;
Gillery, P .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2000, 38 (04) :321-326
[8]   Evaluation of the tosoh HLC-723GHb V A1c 2.2 hemoglobin A1c analyzer [J].
Chevenne, D ;
Marle, N ;
Chauffert, M ;
Noel, M ;
Ducrocq, R ;
Trivin, F .
CLINICAL BIOCHEMISTRY, 1999, 32 (06) :487-490
[9]  
Gibb I, 1999, CLIN CHEM, V45, P1833
[10]  
Goldstein DE, 1982, DIABETES S3, V31, P70