Effects of the bone-plate material and the presence of a gap between the fractured bone and plate on the predicted stresses at the fractured bone

被引:61
作者
Fouad, H. [1 ]
机构
[1] King Saud Univ, RCC, Dept Appl Sci Med, Riyadh 11437, Saudi Arabia
关键词
Bone-plate; Stress shielding; FG plate; SS; Ti; Finite element; FIXATION PLATE; STIFFNESS; INTERFACE;
D O I
10.1016/j.medengphy.2010.05.003
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
In the present study, 3D finite element models for fractured bones with function-graded (FG) bone-plates and traditional bone-plates made of stainless steel (SS) or titanium (Ti) alloy are generated using the ABAOUS code. The predicted Von Mises stresses at the fracture site and underlying bone-plate are examined at different healing stages. The effects on the predicted Von Mises stresses at the fracture site of the presence of a gap between the plate and fractured bone are also studied. Based on the analytical results, it is found that the stress shielding at the fracture site at the fully healed stages decreases when using FG bone-plates compared to Ti alloy or SS bone-plates. In the initial healing stages, the Von Mises stresses at the fracture site increase (stress shielding decreases) by 17% and 11% when using FG bone-plates as compared to SS bone-plates for contacted and non-contacted bone-plate system, respectively. The significant effects of using an FG bone-plate with a gap on the resultant Von Mises stresses on the bone underneath the plate and on the bone stress shielding should be taken into consideration during fractured bone fixation. (C) 2010 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:783 / 789
页数:7
相关论文
共 23 条
[1]
Biomechanical testing of the locking compression plate: When does the distance between bone and implant significantly reduce construct stability? [J].
Ahmad, M. ;
Nanda, R. ;
Bajwa, A. S. ;
Candal-Couto, J. ;
Green, S. ;
Hui, A. C. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2007, 38 (03) :358-364
[2]
CARBON-FIBER COMPOSITE BONE PLATES - DEVELOPMENT, EVALUATION AND EARLY CLINICAL-EXPERIENCE [J].
ALI, MS ;
FRENCH, TA ;
HASTINGS, GW ;
RAE, T ;
RUSHTON, N ;
ROSS, ERS ;
WYNNJONES, CH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (04) :586-591
[3]
Evaluation of bone plate with low-stiffness material in terms of stress distribution [J].
Benli, Semih ;
Aksoy, Sami ;
Havitcioglu, Hasan ;
Kucuk, Mumin .
JOURNAL OF BIOMECHANICS, 2008, 41 (15) :3229-3235
[4]
Discourse concerning two conventional sciences - Prologue [J].
Cordey, J .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 :V-XIV
[5]
Cordey J, 2000, Injury, V31 Suppl 3, pC21
[6]
Biomechanical and histological evaluation of the application of biodegradable poly-L-lactic cushion to the plate internal fixation for bone fracture healing [J].
Fan, Yubo ;
Xiu, Kaihua ;
Duan, Hong ;
Zhang, Ming .
CLINICAL BIOMECHANICS, 2008, 23 (SUPLL.1) :S7-S16
[7]
An evaluation of interface contact profiles in two low contact bone plates [J].
Field, JR ;
Edmonds-Wilson, R ;
Stanley, RM .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2004, 35 (06) :551-556
[8]
Feasibility of knitted carbon/PEEK composites for orthopedic bone plates [J].
Fujihara, K ;
Huang, ZM ;
Ramakrishna, S ;
Satknanantham, K ;
Hamada, H .
BIOMATERIALS, 2004, 25 (17) :3877-3885
[9]
Performance study of braided carbon/PEEK composite compression bone plates [J].
Fujihara, K ;
Huang, ZM ;
Ramakrishna, S ;
Satknanantham, K ;
Hamada, H .
BIOMATERIALS, 2003, 24 (15) :2661-2667
[10]
Biomechanics of bone-fracture fixation by stiffness-graded plates in comparison with stainless-steel plates [J].
Ganesh, V. K. ;
Ramakrishna, K. ;
Ghista, Dhanjoo N. .
BIOMEDICAL ENGINEERING ONLINE, 2005, 4 (1)