An ion-interaction RP-HPLC method for the determination of the eleven EPA priority pollutant phenols

被引:32
作者
Angelino, S
Gennaro, MC
机构
[1] Dipartimento di Chimica Analitica, Univ. degli Studi di Torino, 10125 Torino, via P. Giuria
关键词
high-performance liquid chromatography; priority pollutant phenols; phenols;
D O I
10.1016/S0003-2670(97)00124-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An ion-interaction RP-HPLC method is presented for the simultaneous separation and determination of the eleven EPA priority pollutant phenols. The method is based on the capability of phenolic compounds to form ion-pairs with alkylammonium ions. The optimised mobile phase is a water/acetonitrile solution of octylammonium o-phosphate at pH 8. With UV detection at 285 nm, determination limits lower than 30 mu g l(-1) are achieved, without preconcentration. By the use of statistical tests, the accuracy and the reproducibility of the method is tested (confidence level 95%) with respect to EPA certified standard solution and the method is validated through intercalibration with a GC-MS method. The statistical difference between the methods is evaluated by parametric (F-test, paired t-test, t-test for multiple samples) and non parametric tests (Mann-Whitney, U-test and Wilcoxon test) and the results are discussed. Accuracy, reproducibility and detection limit of the W-HPLC method resulted to be more advantageous with respect to gas-chromatographic methods. Examples of preconcentration and application to the analysis of river water are presented.
引用
收藏
页码:61 / 71
页数:11
相关论文
共 37 条