Cleaning of rotary nickel-titanium endodontic instruments

被引:42
作者
Linsuwanont, P [1 ]
Parashos, P [1 ]
Messer, HH [1 ]
机构
[1] Univ Melbourne, Sch Dent Sci, Melbourne, Vic 3000, Australia
关键词
cleaning; cleaning of NiTi instruments; cross-infection; endodontic instruments; nickel-titanium;
D O I
10.1111/j.1365-2591.2004.00747.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim To develop and evaluate an effective cleaning procedure for rotary nickel-titanium (NiTi) endodontic instruments. Methodology New rotary instruments (ProFile size 25/.04) were contaminated by preparing canals of extracted teeth. Three factors were evaluated to develop an effective cleaning sequence: dry or moist storage before cleaning; mechanical removal (brushing); and chemical dissolution in 1% NaOCl with ultrasonication. Debris on flutes was scored after staining in situ with Van Gieson's solution at x45 magnification. Debris was classified as stained or unstained particulate debris and organic film, and rated as none, slight, moderate or heavy. The effectiveness of a recommended cleaning sequence was tested on different instrument types and in private endodontic practices. Results All new instruments showed metallic spurs and fine particulate debris on the surfaces. After contamination, brushing alone removed most particulate debris, but did not remove organic film. NaOCl effectively removed organic film. Under laboratory conditions, the sequential cleaning procedures (moist storage, brushing followed by immersion in 1% NaOCl and ultrasonic cleaning) totally removed organic debris. Dry storage before cleaning or autoclaving with debris present reduced cleaning effectiveness (P < 0.001, one-way ANOVA). In three private practices, the cleaning protocol substantially reduced biological contamination, but complete cleaning was not always achieved (87% clean). Conclusion Complete removal of organic debris from instruments is feasible using a combination of mechanical removal and chemical dissolution, but requires meticulous attention to details.
引用
收藏
页码:19 / 28
页数:10
相关论文
共 28 条
[1]   ANTIBACTERIAL PROPERTIES OF SODIUM DICHLOROISOCYANURATE AND SODIUM-HYPOCHLORITE FORMULATIONS [J].
BLOOMFIELD, SF ;
MILES, GA .
JOURNAL OF APPLIED BACTERIOLOGY, 1979, 46 (01) :65-73
[2]  
Busslinger A, 1998, INT ENDOD J, V31, P290
[3]  
CARSON LF, 1990, HISTOTECHNOLOGY SELF
[4]   Wear of nickel-titanium lightspeed instruments evaluated by scanning electron microscopy [J].
Eggert, C ;
Peters, O ;
Barbakow, F .
JOURNAL OF ENDODONTICS, 1999, 25 (07) :494-497
[5]   Cyclic fatigue of Profile rotary instruments after prolonged clinical use [J].
Gambarini, G .
INTERNATIONAL ENDODONTIC JOURNAL, 2001, 34 (05) :386-389
[6]   Mechanical properties of nickel-titanium endodontic instruments and the effect of sodium hypochlorite treatment [J].
Haikel, Y ;
Serfaty, R ;
Wilson, P ;
Speisser, JM ;
Allemann, C .
JOURNAL OF ENDODONTICS, 1998, 24 (11) :731-735
[7]  
Hand R E, 1978, J Endod, V4, P60, DOI 10.1016/S0099-2399(78)80255-6
[8]   Cleaning of endodontic files .1. The effect of bioburden on the sterilization of endodontic files [J].
Johnson, MA ;
Primack, PD ;
Loushine, RJ ;
Craft, DW .
JOURNAL OF ENDODONTICS, 1997, 23 (01) :32-34
[9]  
KOSKINEN KP, 1980, SCAND J DENT RES, V88, P406
[10]   A study of blood contamination of Siqveland matrix bands [J].
Lowe, AH ;
Bagg, J ;
Burke, FJT ;
MacKenzie, D ;
McHugh, S .
BRITISH DENTAL JOURNAL, 2002, 192 (01) :43-45