The effect of palladium modifier on the efficiency of antimony hydride trapping in graphite furnace atomic absorption spectrometry (AAS)

被引:8
作者
Garbos, S [1 ]
Bulska, E [1 ]
Hulanicki, A [1 ]
机构
[1] Univ Warsaw, Dept Chem, PL-02093 Warsaw, Poland
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1998年 / 361卷 / 01期
关键词
D O I
10.1007/s002160050831
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The performance of palladium as permanent chemical modifier electroplated on the surface of a graphite tube for the preconcentration of antimony hydride was examined and compared with thermally formed Pd-coatings. The application of Pd-electroplated tubes allows to perform at least 75 determination cycles without any significant change in the efficiency of hydride collection. This is an advantage over the thermally formed palladium coating which must be obtained individually before each measurement. After the optimization of the system parameters a concentration detection limit of 48 ng/L and an absolute detection limit of 71 pg for a 1.48 mi, sample were obtained. The procedure was applied to the determination of antimony at a concentration level of 0.2 mu g/L in a tap water sample.
引用
收藏
页码:43 / 46
页数:4
相关论文
共 25 条
[1]   APPLICATION OF PALLADIUM-PLATING AND RHODIUM-PLATING OF THE GRAPHITE-FURNACE IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
BULSKA, E ;
JEDRAL, W .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (01) :49-53
[2]   DIFFERENT SAMPLE INTRODUCTION SYSTEMS FOR THE SIMULTANEOUS DETERMINATION OF AS, SB AND SE BY MICROWAVE-INDUCED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
BULSKA, E ;
TSCHOPEL, P ;
BROEKAERT, JAC ;
TOLG, G .
ANALYTICA CHIMICA ACTA, 1993, 271 (01) :171-181
[3]   Comparison of chemical modifiers for the determination of selenium by electrothermal atomic-absorption spectrometry [J].
Bulska, E ;
Pyrzynska, K .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (9-10) :1283-1291
[4]   Noble metals as permanent modifiers for the determination of mercury by electrothermal atomic absorption spectrometry [J].
Bulska, E ;
Kandler, W ;
Hulanicki, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (9-10) :1263-1270
[5]  
DOIDGE PS, 1989, J ANAL ATOM SPECTROM, V4, P251, DOI 10.1039/ja9890400251
[6]   APPLICATION OF THE FURNACE ATOMIC-ABSORPTION METHOD FOR THE DETECTION OF ARSENIC IN BIOLOGICAL SAMPLES BY MEANS OF THE HYDRIDE TECHNIQUE [J].
DRASCH, G ;
MEYER, LV ;
KAUERT, G .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1980, 304 (2-3) :141-142
[7]   Preconcentration of inorganic species of antimony by sorption on Polyorgs 31 followed by atomic absorption spectrometry detection [J].
Garbos, S ;
Bulska, E ;
Hulanicki, A ;
Shcherbinina, NI ;
Sedykh, EM .
ANALYTICA CHIMICA ACTA, 1997, 342 (2-3) :167-174
[8]  
GARBOS S, 1995, SPECTROCHIM ACTA B, V50, P1673
[9]   Investigations into the role of modifiers for entrapment of hydrides in flow injection hydride generation electrothermal atomic absorption spectrometry as exemplified by the determination of Germanium [J].
Hilligsoe, B ;
Andersen, JET ;
Hansen, EH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (05) :585-588
[10]   INVESTIGATION OF HIGH-TEMPERATURE REACTIONS ON SOLID SUBSTRATES WITH RUTHERFORD BACKSCATTERING SPECTROMETRY - INTERACTION OF PALLADIUM WITH SELENIUM ON HEATED GRAPHITE SURFACES [J].
MAJIDI, V ;
ROBERTSON, JD .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (13) :1723-1733