Compound retention dependence of the response in a gas chromatography atomic emission detection system

被引:8
作者
Augusto, F [1 ]
Valente, ALP [1 ]
机构
[1] Univ Estadual Campinas, Inst Chem, UNICAMP, BR-13083970 Sao Paulo, Brazil
关键词
microwave-induced helium plasma atomic emission detection; detection; GC; retention-detection response relationships; alkanes; arenes; haloarenes;
D O I
10.1016/S0021-9673(98)00483-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The relationship between the response of a laboratory-made GC-microwave-induced helium plasma atomic emission detection (MIP) system prototype and compound nature was studied by monitoring the carbon 247.9 nm atomic emission line. The test samples for this study were solutions of alkanes, arenes and haloarenes prepared at two concentration levels; varied plasma generation conditions (microwave input powers and helium support plasma gas flows) were evaluated. The statistical examination of the data pointed to a dependence between MIP signal and compound retention - less retained compounds tend to show reduced MIP signals. (C) 1998 Elsevier Science BN. All rights reserved.
引用
收藏
页码:85 / 91
页数:7
相关论文
共 16 条
[1]  
BAJPAI AC, 1978, STAT METHODS ENG SCI, P195
[2]  
BEENAKKER CIM, 1980, PHILIPS TECH REV, V39, P65
[3]  
Bell A. T., 1974, Techniques and applications of plasma chemistry, P1
[4]  
GROB RL, 1977, MODERN PRACTICE GAS, P51
[5]   MOLECULAR FORMULA DETERMINATION OF CAPILLARY COLUMN GAS-CHROMATOGRAPHIC EFFLUENTS USING COMBINED MICROWAVE-INDUCED PLASMA EMISSION AND MASS-SPECTRAL DATA [J].
HOOKER, DB ;
DEZWAAN, J .
ANALYTICAL CHEMISTRY, 1989, 61 (19) :2207-2211
[6]  
HUANG YR, 1990, J ANAL ATOM SPECTROM, V5, P115
[7]  
KINGJAN HA, 1983, SPECTROCHIM ACTA B, V38, P777
[8]  
KINGJAN HA, 1981, SPECTROCHIM ACTA B, V36, P325
[9]  
MILLER JC, 1993, ESTADISTICA QUIMICA, P42
[10]  
PEDERSENBJERGAARD S, 1992, HRC-J HIGH RES CHROM, V15, P89