Surface Topographies of Glaucoma Drainage Devices and Their Influence on Human Tenon Fibroblast Adhesion

被引:42
作者
Choritz, Lars [1 ]
Koynov, Kaloian [2 ]
Renieri, Giulia [1 ]
Barton, Keith [3 ]
Pfeiffer, Norbert [1 ]
Thieme, Hagen [1 ]
机构
[1] Johannes Gutenberg Univ Mainz, Dept Ophthalmol, Univ Med Ctr, BiomaT Res Grp, D-55131 Mainz, Germany
[2] Max Planck Inst Polymer Res, D-55128 Mainz, Germany
[3] Moorfields Eye Hosp, London, England
关键词
RANDOMIZED CLINICAL-TRIAL; 500-MM(2) BAERVELDT IMPLANT; PLATE MOLTENO IMPLANTATION; ADJUNCTIVE MITOMYCIN-C; CYCLIC TENSILE STRAIN; AHMED GLAUCOMA; VALVE IMPLANT; REFRACTORY GLAUCOMA; DIFFERENT BIOMATERIALS; PEDIATRIC-PATIENTS;
D O I
10.1167/iovs.09-4759
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
PURPOSE. This study was performed to investigate the surface topography of different glaucoma drainage devices and to determine the effects of surface roughness on cell adhesion of cultured human tenon fibroblasts. METHODS. The surface topography of four widely used devices (Ahmed FP7 and Ahmed S-2; New World Medical, Inc., Rancho Cucamonga, CA; Baerveldt BG101-350; Advanced Medical Optics, Irvine, CA; and Molteno S1; Molteno Ophthalmic Ltd., Dunedin, New Zealand) was investigated by scanning electron microscopy, and roughness was quantified by white-light confocal microscopy. Cells were grown for 72 hours on the surfaces of implants affixed to standard culture dishes. The cells were labeled with a fluorescent dye and detected by confocal laser scanning microscopy, while simultaneously imaging the surface reflectance. Collagen adsorption was quantified immunologically by using fluorescent beads coupled to a secondary antibody. RESULTS. The root-mean-square roughness was 1.5 +/- 0.1 mu m (mean +/- SE) for the silicone Ahmed model FP7 and 1.3 +/- 0.1 mu m for the Ahmed with polypropylene base plate Ahmed model S-2. The Baerveldt was substantially smoother, with a mean roughness of 0.1 +/- 0.01 mu m. The Molteno was the smoothest of all devices (0.07 +/- 0.01 mu m). Cell adhesion was most prevalent on base plates with higher surface roughness, markedly less pronounced on the smoother base plates, and independent of collagen adsorption. CONCLUSIONS. The most frequently implanted glaucoma drainage devices are of markedly different surface topography. Surface roughness appears to correlate with tenon fibroblast adhesion in vitro and also with the rate of occurrence of postimplantation hypertensive phase and failure due to fibrous encapsulation. Surface roughness may thus play a role in triggering excessive fibrovascular reactions. Smoother base plate surfaces may enhance the success rates of these devices. (Invest Ophthalmol Vis Sci. 2010;51:4047-4053) DOI:10.1167/iovs.09-4759
引用
收藏
页码:4047 / 4053
页数:7
相关论文
共 46 条
[1]
Glaucoma drainage implants in the treatment of refractory glaucoma in pediatric patients [J].
Autrata, R. ;
Helmanova, I. ;
Oslejskova, H. ;
Vondracek, P. ;
Rehurek, J. .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2007, 17 (06) :928-937
[2]
A clinical study of the Ahmed glaucoma valve implant in advanced glaucoma [J].
Ayyala, RS ;
Zurakowski, D ;
Smith, JA ;
Monshizadeh, R ;
Netland, PA ;
Richards, DW ;
Layden, WE .
OPHTHALMOLOGY, 1998, 105 (10) :1968-1976
[3]
Ayyala RS, 2002, OPHTHALMIC SURG LAS, V33, P94
[4]
Comparison of different biomaterials for glaucoma drainage devices - Part 2 [J].
Ayyala, RS ;
Michelini-Norris, B ;
Flores, A ;
Haller, E ;
Margo, CE .
ARCHIVES OF OPHTHALMOLOGY, 2000, 118 (08) :1081-1084
[5]
Ayyala RS, 1999, ARCH OPHTHALMOL-CHIC, V117, P233
[6]
Randomized clinical trial of the 350-mm2 versus the 500-mm2 Baerveldt implant:: Longer term results -: Is bigger better? [J].
Britt, MT ;
LaBree, LD ;
Lloyd, MA ;
Minckler, DS ;
Heuer, DK ;
Baerveldt, G ;
Varma, R .
OPHTHALMOLOGY, 1999, 106 (12) :2312-2318
[7]
Cantor L, 1998, J GLAUCOMA, V7, P240
[8]
Induction of tenascin-C by cyclic tensile strain versus growth factors:: distinct contributions by Rho/ROCK and MAPK signaling pathways [J].
Chiquet, M ;
Sarasa-Renedo, A ;
Tunç-Civelek, V .
BIOCHIMICA ET BIOPHYSICA ACTA-MOLECULAR CELL RESEARCH, 2004, 1693 (03) :193-204
[9]
From mechanotransduction to extracellular matrix gene expression in fibroblasts [J].
Chiquet, Matthias ;
Gelman, Laurent ;
Lutz, Roman ;
Maier, Silke .
BIOCHIMICA ET BIOPHYSICA ACTA-MOLECULAR CELL RESEARCH, 2009, 1793 (05) :911-920
[10]
Histopathologic and immunohistochemical analysis of the filtration bleb after unsuccessful glaucoma seton implantation [J].
Classen, L ;
Kivela, T ;
Tarkkanen, A .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1996, 122 (02) :205-212