Spatial resolution in Micromegas detectors

被引:33
作者
Derré, J [1 ]
Giomataris, Y
Zaccone, H
Bay, A
Perroud, JP
Ronga, F
机构
[1] CEA Saclay, DAPNIA, F-91191 Gif Sur Yvette, France
[2] Univ Lausanne, IPHE, BSP, CH-1015 Lausanne, Switzerland
关键词
Micromegas detector; charged particle tracking in gaseous detector; spatial resolution; spatial accuracy; space resolution; space accuracy;
D O I
10.1016/S0168-9002(00)01051-2
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The performance of a telescope of Micromegas detectors has been studied in a pion beam at the CERN PS. With a gas filling of CF4 and 20% isobutane and with a strip pitch of 100 mum an accuracy of 14 +/- 3 mum on the spatial resolution has been measured at normal incidence. A simulation demonstrates that the resolution is limited by the size of the holes of the mesh of the detector and could be reduced to 11 mum in the same conditions with smaller holes. Even further improvement down to 8.5 mum is feasible for the same gas with an optimized 75 mum strip pitch. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:523 / 531
页数:9
相关论文
共 2 条
[1]   MICROMEGAS: A high-granularity position-sensitive gaseous detector for high particle-flux environments [J].
Giomataris, Y ;
Rebourgeard, P ;
Robert, JP ;
Charpak, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 376 (01) :29-35
[2]   MEASUREMENT OF ELECTRON-DRIFT PARAMETERS FOR HELIUM AND CF4-BASED GASES [J].
VAVRA, J ;
COYLE, P ;
KADYK, J ;
WISE, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 324 (1-2) :113-126