Comparison of thermal tissue effects induced by contact application of fiber guided laser systems

被引:59
作者
Janda, P
Sroka, R
Mundweil, B
Betz, CS
Baumgartner, R
Leunig, A
机构
[1] Univ Munich, Klinikum Grosshadern, Dept Otorhinolaryngol Head & Neck Surg, D-81377 Munich, Germany
[2] Univ Munich, Laser Res Lab, Dept Urol, D-81377 Munich, Germany
关键词
contact laser treatment; diode-laser; Ho : YAG-laser; Nd : YAG-laser; thermal laser effects;
D O I
10.1002/lsm.10199
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Background and Objectives: Laser light of various wavelengths is being used for surgical procedures in otolaryngology. Apart from well-known fiber guided laser systems such as Nd:YAG- and Ho:YAG-lasers, newly developed diode-laser systems of different wavelengths have recently become popular in surgery. In order to compare the effects of fiber guided laser light with respect to their induced tissue effects, these laser systems have been studied and compared under reproducible test conditions. Study Design/Materials and Methods: The laser fibers of four common medical laser systems (Ho:YAG- (lambda= 2,080 nm), Nd:YAG- (lambda = 1,064 nm), and diode-laser (lambda = 830 and 940 nm)) were fixed to a computer controlled stepper motor. The laser light was applied in contact mode onto ex vivo muscle tissue, using identical power settings and a reproducible application procedure (application velocity, application angle) under constant conditions (temperature of tissue and volume). The size of the thermal effects on the tissue (e.g., coagulation, ablation, and carbonization zones) were measured and photographed via optical microscopy. Results: Depending on the laser wavelength used, the experimental results proved different degrees of tissue responses. Nd:YAG- and diode-lasers provided for only low coagulation effects in the depth of the tissue, but produced severe carbonization at the surface. Ho:YAG-laser light revealed the highest ablation capabilities of the lasers investigated in addition to large coagulation zones which were of larger extent than those produced by Nd:YAG- and diode-laser light. Conclusions: Contact treatment by Ho:YAG-laser light might provide for a precise and effective tissue reduction in a bloodless manner because of its high ablation and coagulation capabilities, especially if large volumes are treated and structures beneath are non-critical. In comparison, Nd:YAG- and diode-laser treatment in contact application showed low thermal tissue effects (i.e., coagulation) in the depth, resulting from a high power loss caused by the development of large carbonization zones at the surface of the tissue. Therefore, the degree of blood-perfusion and the capability of vessel-closure induced by these lasers should be taken into account. The presented investigation also revealed that in contact mode, the tested laser systems produced tissue effects, which were highly different from those already described for applications in non-contact mode. Physicians who are performing laser treatments in close boundaries must be aware that changing from non-contact to contact mode in laser application greatly influences the resulting tissue effects. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:93 / 101
页数:9
相关论文
共 31 条
[1]   Long-term results of inferior turbinate reduction with argon plasma coagulation [J].
Bergler, WF ;
Sadick, H ;
Hammerschmitt, N ;
Oulmi, J ;
Hörmann, K .
LARYNGOSCOPE, 2001, 111 (09) :1593-1598
[2]   Transoral CO2 laser for surgical management of glottic carcinoma in situ [J].
Damm, M ;
Sittel, C ;
Streppel, M ;
Eckel, HE .
LARYNGOSCOPE, 2000, 110 (07) :1215-1221
[3]  
Densert O, 2001, ACTA OTO-LARYNGOL, V121, P854
[4]   Subjective comparison of Nd:YAG, diode, and CO2 lasers for endoscopically guided inferior turbinate reduction surgery [J].
DeRowe, A ;
Landsberg, R ;
Leonov, Y ;
Katzir, A ;
Ophir, D .
AMERICAN JOURNAL OF RHINOLOGY, 1998, 12 (03) :209-212
[5]  
FEYH J, 1995, ADV OTO-RHINO-LARYNG, V49, P122
[6]  
GLEICH LL, 1995, ARCH OTOLARYNGOL, V121, P1162
[7]  
GUO J, 1998, CHIN MED J ENGL, V111, P928
[8]   Argon plasma coagulation (APC) in palliative surgery of head and neck malignancies [J].
Hauser, U ;
Hoffmann, TK ;
Balló, H ;
Bier, H .
LARYNGOSCOPE, 2002, 112 (07) :1275-1280
[9]  
Helfmann J., 1989, ANGEW LASERMEDIZIN L, P1
[10]   Clinical experience with power-regulated contact laser surgery for the paranasal sinuses and the anterior skull base [J].
Ilgner, J ;
Emmerling, O ;
Biesterfeld, S ;
Westhofen, M .
LARYNGO-RHINO-OTOLOGIE, 2002, 81 (05) :346-350