Percutaneous coronary interventions using a new 5 french guiding catheter:: Results of a prospective study

被引:13
作者
Schöbel, WA [1 ]
Spyridopoulos, I [1 ]
Hoffmeister, HM [1 ]
Seipel, L [1 ]
机构
[1] Univ Tubingen, Dept Cardiol, D-72076 Tubingen, Germany
关键词
guiding catheter; 5; French; percutaneous coronary intervention;
D O I
10.1002/ccd.1172
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this prospective study was to analyze the technical feasibility, the success rate, and the special complications of percutaneous coronary interventions (PCIs) using a newly released 5 Fr guiding catheter with an inner diameter of 0.058 ". The study was performed in 150 consecutive patients subjected to coronary angioplasty, In 89% of the patients, the intervention was started with a 5 Fr catheter (JR4 or JL4); in 16 patients a 6 or 7 Fr catheter was used because of unstable clinical conditions according to the decision of the interventional cardiologist. In 12 out of 134 patients, the guiding catheter had to be changed during the intervention from 5 Fr to a 6 or 7 Fr catheter due to poor backup support. In 112 out of 118 patients, the intervention was successfully performed using a 5 Fr catheter (95%); in 12 out of 16 patients, after changing the guiding catheter, the overall success rate was 93%. In patients with type A and B lesions who were initially treated using a 5 Fr catheter, the procedural success rate was 100% (81 out of 81), whereas in patients with type C lesions the procedural success rate was 83% (43 out of 53; P = 0.000053, Fisher's exact test). Furthermore, in patients with a diameter stenosis < 90%, the procedural success rate was 100% (57 out of 57), whereas in patients with a diameter stenosis of 90%-100%, the procedural success rate was 87% (67 out of 77; P = 0.0050). Stent implantation was performed successfully in 24 patients (18%) using the 5 Fr guiding catheter, This study confirms that PCI was technically feasible using a 5 Fr guiding catheter in the majority of consecutive patients with a success rate of 95%. There were significant differences in the success rate depending on the lesion type and the diameter stenosis. Complications were very rare and were not related to the guiding catheter, Limitations of the 5 Fr guiding catheters arose mainly from a poor backup support in long lesions and severe stenosis, (C) 2001 Wiley-Liss. Inc.
引用
收藏
页码:308 / 312
页数:5
相关论文
共 17 条
[1]   EVALUATION OF THE SYSTEMATIC USE OF A NEW-TYPE OF DIAGNOSTIC INTERVENTIONAL 6F CATHETERS FOR CORONARY ANGIOGRAPHY AND ANGIOPLASTY [J].
CHATELAIN, P ;
URBAN, P ;
CAMENZIND, E ;
VERINE, V ;
HOANG, V ;
METZ, D .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1994, 31 (04) :345-349
[2]  
Gobeil F, 2000, CATHETER CARDIO INTE, V51, P107, DOI 10.1002/1522-726X(200009)51:1<107::AID-CCD24>3.0.CO
[3]  
2-L
[4]   Intracoronary stent implantation without ultrasound guidance and with replacement of conventional anticoagulation by antiplatelet therapy - 30-day clinical outcome of the French Multicenter Registry [J].
Karrillon, GJ ;
Morice, MC ;
Benveniste, E ;
Bunouf, P ;
Aubry, P ;
Cattan, S ;
Chevalier, B ;
Commeau, P ;
Cribier, A ;
Eiferman, C ;
Grollier, G ;
Guerin, Y ;
Henry, M ;
Lefevre, T ;
Livarek, B ;
Louvard, Y ;
Marco, J ;
Makowski, S ;
Monassier, JP ;
Pernes, JM ;
Rioux, P ;
Spaulding, C ;
Zemour, G .
CIRCULATION, 1996, 94 (07) :1519-1527
[5]   Nonrandomized comparison between stent deployment and percutaneous transluminal coronary angioplasty in acute myocardial infarction [J].
Katz, S ;
Green, SJ ;
Kaplan, BM ;
Chepurko, L ;
Goldner, BG ;
Ong, L .
AMERICAN HEART JOURNAL, 2000, 139 (01) :44-51
[6]   Comparison of 6F with 7F and 8F guiding catheters for elective coronary angioplasty: Results of a prospective, multicenter, randomised trial [J].
Metz, D ;
Meyer, P ;
Touati, C ;
Coste, P ;
Petiteau, PY ;
Durand, P ;
Faivre, R ;
Lefevre, T ;
Elaerts, J .
AMERICAN HEART JOURNAL, 1997, 134 (01) :131-137
[7]  
MORISDE IT, 1996, REV ESP CARDIOL, V49, P124
[8]  
Noguchi T, 2000, CATHETER CARDIO INTE, V49, P258, DOI 10.1002/(SICI)1522-726X(200003)49:3<258::AID-CCD7>3.3.CO
[9]  
2-C
[10]  
Peels HOJ, 1996, CATHETER CARDIO DIAG, V38, P210, DOI 10.1002/(SICI)1097-0304(199606)38:2<210::AID-CCD21>3.0.CO