Knee cartilage topography, thickness, and contact areas from MRI: in-vitro calibration and in-vivo measurements

被引:293
作者
Cohen, ZA
McCarthy, DM
Kwak, SD
Legrand, P
Fogarasi, F
Ciaccio, EJ
Ateshian, GA
机构
[1] Columbia Univ, Orthopaed Res Lab, Dept Mech Engn, New York, NY 10032 USA
[2] Columbia Univ, Dept Orthopaed Surg, New York, NY 10032 USA
[3] Columbia Univ, Dept Radiol, New York, NY 10032 USA
关键词
MRI accuracy; knee surface topography; cartilage thickness; contact area;
D O I
10.1053/joca.1998.0165
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: This study assessed the three-dimensional accuracy of magnetic resonance imaging (MRI) for measuring articular surface topographies and cartilage thicknesses of human cadaveric knee joints, by comparison with the calibrated stereophotogrammetric (SPG) method. Methods: Six fresh frozen cadaveric knees and the knees of four volunteers were imaged with a three-dimensional spoiled gradient-recalled acquisition with fat suppression using a linear extremity coil in a 1.5 T superconducting magnet. The imaging voxel size was 0.47 x 0.47 x 1.0 mm. Both a manual and a semi-automated segmentation method were employed to extract topographic measurements from MRI. Following MRI, each of the six cadaveric knees was dissected and its articular surfaces quantified using stereophotogrammetry. The MRI surface measurements were compared numerically with the SPG; measurements. Results: For six cadaveric knees, the average accuracies of cartilage and subchondral bone surface measurements were found to be 0.22 mm and 0.14 mm respectively and the thickness measurements demonstrated an average accuracy of 0.31 mm. It was found that while most of the error may be attributed to random measurement error, the accuracy was somewhat affected by systematic errors. For each bone of the knee, accuracies were most favorable in the patella, followed by the femur and then the tibia. The more efficient semi-automated method provided equally good and sometimes better accuracies than manual segmentation. Conclusions: This study demonstrates that clinical MRI can provide accurate measurements of cartilage topography, thickness, contact areas and surface curvatures of the knee.
引用
收藏
页码:95 / 109
页数:15
相关论文
共 38 条
[1]   INVITRO MEASUREMENT OF STATIC PRESSURE DISTRIBUTION IN SYNOVIAL JOINTS .2. RETROPATELLAR SURFACE [J].
AHMED, AM ;
BURKE, DL ;
YU, A .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1983, 105 (03) :226-234
[2]  
[Anonymous], NUMERICAL RECIPES FO
[3]   CURVATURE CHARACTERISTICS AND CONGRUENCE OF THE THUMB CARPOMETACARPAL JOINT - DIFFERENCES BETWEEN FEMALE AND MALE JOINTS [J].
ATESHIAN, GA ;
ROSENWASSER, MP ;
MOW, VC .
JOURNAL OF BIOMECHANICS, 1992, 25 (06) :591-+
[4]   A STEREOPHOTOGRAMMETRIC METHOD FOR DETERMINING IN-SITU CONTACT AREAS IN DIARTHROIDAL JOINTS, AND A COMPARISON WITH OTHER METHODS [J].
ATESHIAN, GA ;
KWAK, SD ;
SOSLOWSKY, LJ ;
MOW, VC .
JOURNAL OF BIOMECHANICS, 1994, 27 (01) :111-124
[5]   A B-SPLINE LEAST-SQUARES SURFACE-FITTING METHOD FOR ARTICULAR SURFACES OF DIARTHROIDAL JOINTS [J].
ATESHIAN, GA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (04) :366-373
[6]   QUANTITATION OF ARTICULAR SURFACE-TOPOGRAPHY AND CARTILAGE THICKNESS IN KNEE JOINTS USING STEREOPHOTOGRAMMETRY [J].
ATESHIAN, GA ;
SOSLOWSKY, LJ ;
MOW, VC .
JOURNAL OF BIOMECHANICS, 1991, 24 (08) :761-776
[7]  
ATESHIAN GA, 1995, ASME ADV BIOENG, V31, P225
[8]   KNEE HYALINE CARTILAGE EVALUATED WITH MR IMAGING - A CADAVERIC STUDY INVOLVING MULTIPLE IMAGING SEQUENCES AND INTRAARTICULAR INJECTION OF GADOLINIUM AND SALINE SOLUTION [J].
CHANDNANI, VP ;
HO, C ;
CHU, P ;
TRUDELL, D ;
RESNICK, D .
RADIOLOGY, 1991, 178 (02) :557-561
[9]  
COHEN ZA, 1997, T ORTHOP RES SOC, V22, P625
[10]  
COHEN ZA, 1997, ASME BED, V35, P45