Frequent ICD shocks due to double sensing in patients with bi-ventricular implantable cardioverter defibrillators

被引:10
作者
Al-Ahmad, A [1 ]
Wang, PJ [1 ]
Homoud, MK [1 ]
Estes, NAM [1 ]
Link, MS [1 ]
机构
[1] Tufts New England Med Ctr, Dept Med, Div Cardiol, New England Cardiac Arrhythmia Ctr, Boston, MA 02111 USA
关键词
implantable cardioverter defibrillator; biventricular pacing; heart failure;
D O I
10.1023/A:1027455713120
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Biventricular pacing has emerged as a modality for treatment of patients with heart failure. Combined biventricular pacers and implantable cardioverter defibrillators offer treatment of heart failure as well as protection from sudden cardiac death. However, inappropriate ICD shocks as a result of double sensing due to widely spaced ventricular bipoles may pose a significant problem in these patients. We examined the ICD records of twenty-three patients with biventricular ICDs, and evaluated all episodes of double sensing that resulted in aborted or delivered therapy. In follow-up of 3.7 +/- 2.6 months, thirty-three shocks in fifteen episodes occurred in five patients (21.7%) due to double sensing. Four patients (17.4%) had aborted shocks due to double sensing. All episodes resulting in shock occurred because of sinus tachycardia or supraventricular tachycardia above the upper programmed pacing rate of the device with resultant AV conduction and double sensing of the nonpaced ventricular depolarization. In conclusion, double sensing of the R-wave is a common and clinically important cause of inappropriate ICD detection and shock in patients with biventricular ICDs. Appropriate programming of the ICD can prevent episodes of inappropriate shocks.
引用
收藏
页码:377 / 381
页数:5
相关论文
共 17 条
[1]  
Bakker PF, 2000, J INTERV CARD ELECTR, V4, P395
[2]   Inappropriate shock therapy in a heart failure defibrillator [J].
Betts, TR ;
Allen, S ;
Roberts, PR ;
Morgan, JM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (02) :238-240
[3]  
Bocchiardo M, 2000, PACE, V23, P1989
[4]   Reduction of hospital days by biventricular pacing [J].
Braunschweig, F ;
Linde, C ;
Gadler, F ;
Rydén, L .
EUROPEAN JOURNAL OF HEART FAILURE, 2000, 2 (04) :399-406
[5]   A randomized study of the prevention of sudden death in patients with coronary artery disease [J].
Buxton, AE ;
Lee, KL ;
Fisher, JD ;
Josephson, ME ;
Prystowsky, EN ;
Hafley, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1999, 341 (25) :1882-1890
[6]   Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. [J].
Cazeau, S ;
Leclercq, C ;
Lavergne, T ;
Walker, S ;
Varma, C ;
Linde, C ;
Garrigue, S ;
Kappenberger, L ;
Haywood, GA ;
Santini, M ;
Bailleul, C ;
Daubert, JC .
NEW ENGLAND JOURNAL OF MEDICINE, 2001, 344 (12) :873-880
[7]  
Conraads V, 2001, NEW ENGL J MED, V345, P293
[8]   The acute effects of transvenous biventricular pacing in a patient with congestive heart failure [J].
Debrunner, M ;
Naegeli, B ;
Bertel, O .
CHEST, 2000, 117 (06) :1798-1800
[9]   Should stimulation therapy for congestive heart failure be combined with defibrillation backup? [J].
Gaita, F ;
Bocchiardo, M ;
Porciani, MC ;
Vivalda, L ;
Colella, A ;
Di Donno, P ;
Caponi, D ;
Bruzzone, M ;
Padeletti, L .
AMERICAN JOURNAL OF CARDIOLOGY, 2000, 86 (9A) :165K-168K
[10]   Matching approved "nondedicated" hardware to obtain biventricular pacing and defibrillation: Feasibility and troubleshooting [J].
Kanagaratnam, L ;
Pavia, S ;
Schweikert, R ;
Marrouche, N ;
Lam, C ;
Abreu, M ;
Ching, E ;
Chung, M ;
Saliba, W ;
Niebauer, M ;
Wilkoff, B ;
Tchou, P ;
Natale, A .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (07) :1066-1071