摘要
A 71-year-old man nifh a history of myocardial infarction and syncopal ventricular tachycardia received an ICD. Progressively, increased defibrillation charge times of the ICD were observed. During a fast ventricular tachycardia, the patient collapsed. The underlying malfunction could be documented by ICD telemetric interrogation. An excessive charge time rt as recorded causing subsequent delayed therapy. The ICD was considered potentially ineffective and required a generator exchange. This case report underlines the possible factors affecting the ICD charge times. Excessive charge time must be considered as a rare, or underestimated, cause of ICD malfunction, requiring generator replacement.
引用
收藏
页码:518 / 519
页数:2
相关论文
共 9 条
  • [1] Adams Ted P., 1996, P421
  • [2] Ennis Joel B., 1996, P223
  • [3] COMPLICATIONS OF IMPLANTABLE CARDIOVERTER DEFIBRILLATOR THERAPY - FOLLOW-UP OF 241 PATIENTS
    GRIMM, W
    FLORES, BF
    MARCHLINSKI, FE
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (01): : 218 - 222
  • [4] JORDAENS L, 1996, ADV CARDIOLOGY IMPLA, P81
  • [5] Increased defibrillation threshold due to ventricular fibrillation duration - Potential mechanisms
    Lerman, BB
    Engelstein, ED
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 1995, 28 : 21 - 24
  • [6] Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia
    Moss, AJ
    Hall, WJ
    Cannom, DS
    Daubert, JP
    Higgins, SL
    Klein, H
    Levine, JH
    Saksena, S
    Waldo, AL
    Wilber, D
    Brown, MW
    Heo, M
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1996, 335 (26) : 1933 - 1940
  • [7] Rosenthal Mark E., 1996, P435
  • [8] WINCKLE R, 1990, CIRCULATION, V81, P1477
  • [9] INFLUENCE OF TIME AND THERAPY ON VENTRICULAR DEFIBRILLATION IN DOGS
    YAKAITIS, RW
    EWY, GA
    OTTO, CW
    TAREN, DL
    MOON, TE
    [J]. CRITICAL CARE MEDICINE, 1980, 8 (03) : 157 - 163