Corrosion at the Stem-Sleeve Interface of a Modular Titanium Alloy Femoral Component as a Reason for Impaired Disengagement

被引:26
作者
Fraitzl, Christian R. [1 ]
Moya, Luis E. [1 ]
Castellani, Lorenzo [1 ]
Wright, Timothy M. [2 ]
Buly, Robert L. [1 ]
机构
[1] Hosp Special Surg, Hip & Knee Serv, New York, NY 10021 USA
[2] Hosp Special Surg, Dept Biomech, New York, NY 10021 USA
关键词
total hip replacement; modular femoral component; titanium-alloy; corrosion; TOTAL HIP-ARTHROPLASTY; PROSTHESIS; SYSTEM; WEAR;
D O I
10.1016/j.arth.2009.10.018
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Modularity in sleeved femoral components allows the exchange of the stem without disruption of the fixation between the sleeve and the surrounding bone at revision surgery. Failure to disengage the stem from the sleeve would represent an unnecessary compromise from the intended usefulness of the modular design. We report the results of an examination of 22 modular titanium alloy femoral components retrieved after 0.0 to 8.8 years in vivo. In 7 implants, the stem-sleeve interface could not be disengaged without cutting through the components or using mechanical force. Moderate to severe corrosion was detected in all 7 of these cases. Corrosive surface changes were observed in an additional 6 interfaces. There was no correlation with the length of time that the devices had been implanted. When only the stem is to be revised, orthopedic surgeons should be aware of difficulties in disengagement and anticipate alternative surgical procedures.
引用
收藏
页码:113 / 119
页数:7
相关论文
共 15 条
[1]
BOBYN JD, 1994, CLIN ORTHOP RELAT R, P27
[2]
BOBYN JD, 1993, BRIS MYER Z, P287
[3]
Buly R, 2005, ORTHOPEDICS, V28, pS1025
[4]
The role of modularity in primary total hip arthroplasty [J].
Cameron, Hugh U. ;
Keppler, Louis ;
McTighe, Tim .
JOURNAL OF ARTHROPLASTY, 2006, 21 (04) :89-92
[5]
CHANDLER HP, 1995, CLIN ORTHOP RELAT R, P130
[6]
Primary total hip arthroplasty with use of the modular S-ROM prosthesis - Four to seven-year clinical and radiographic results [J].
Christie, MJ ;
DeBoer, DK ;
Trick, LW ;
Brothers, JC ;
Jones, RE ;
Vise, GT ;
Gruen, TA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (12) :1707-1716
[7]
COOK SD, 1994, CLIN ORTHOP RELAT R, P80
[8]
Cook Stephen D., 1993, Clinical Materials, V12, P153, DOI 10.1016/0267-6605(93)90067-H
[9]
Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149
[10]
FATIGUE BEHAVIOR OF A TITANIUM FEMORAL HIP-PROSTHESIS WITH PROXIMAL SLEEVE-STEM MODULARITY [J].
KRYGIER, JJ ;
DUJOVNE, AR ;
BOBYN, JD .
JOURNAL OF APPLIED BIOMATERIALS, 1994, 5 (03) :195-201