Automated determination of phenylcarbamate herbicides in environmental waters by on-line trace enrichment and reversed-phase liquid chromatography diode array detection

被引:15
作者
Hidalgo, C [1 ]
Sancho, JV [1 ]
López, F [1 ]
Hernández, F [1 ]
机构
[1] Univ Jaume 1, Expt Sci Dept, Castellon de La Plana 12080, Spain
关键词
environmental analysis; water analysis; sample handling; pesticides; phenylcarbamates;
D O I
10.1016/S0021-9673(98)00541-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A fully automated liquid chromatographic method using on-line trace-enrichment, gradient elution and diode array detection is described for the trace-level determination of several phenylcarbamate herbicides, such as carbetamide, propham, desmedipham, phenmedipham, chlorbufam and chlorpropham, in environmental water samples. In this work, two different enrichment pre-columns have been assayed, a 5.8x4.6 mm I.D., 10 mu m ODS Prelute cartridge and a 10x2 mm I.D. cartridge filled with 10 mu m PRP-1 polymer, both coupled to a 150x4.6 mm I.D. analytical column filled with 5 mu m ODS. Using the C-18 pre-column, up to 50 ml of water sample could be percolated without peak broadening of any compound. However, a lack of reproducibility was observed in the case of carbetamide, the most polar analyte, after performing recovery experiments by percolating drinking and surface water samples spiked at several levels (0.5 and 4 mu g l(-1))(.) On the other hand, the PRP-1 pre-column allowed the enrichment up to 100 ml of water sample with satisfactory results for every compound, including carbetamide. The procedure was validated by recovery experiments in environmental water samples spiked at 0.2 and 1 mu g l(-1) yielding average recoveries between 84-108% with relative standard deviations in the range 2-12%. Detection limits as low as 0.04 mu g l(-1) were achieved. It was observed that desmedipham and phenmedipham degraded rapidly in the environmental water samples as showed the degradation studies performed along 24 h in drinking and surface waters spiked at 4 mu g l(-1). Although the standard mixture, prepared in HPLC water, was stable for around one week, in the environmental water matrices more than 95% of each herbicide degraded after 6 h, and new chromatographic peaks corresponding to the degradation products were detected. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:121 / 128
页数:8
相关论文
共 20 条
[1]   ENVIRONMENTAL-PROTECTION-AGENCY AND OTHER METHODS FOR THE DETERMINATION OF PRIORITY PESTICIDES AND THEIR TRANSFORMATION PRODUCTS IN WATER [J].
BARCELO, D .
JOURNAL OF CHROMATOGRAPHY, 1993, 643 (1-2) :117-143
[2]  
Borek V., 1986, Agrochemia, V26, P118
[3]   MULTIRESIDUE METHOD FOR PESTICIDES IN DRINKING-WATER USING A GRAPHITIZED CARBON-BLACK CARTRIDGE EXTRACTION AND LIQUID-CHROMATOGRAPHIC ANALYSIS [J].
DICORCIA, A ;
MARCHETTI, M .
ANALYTICAL CHEMISTRY, 1991, 63 (06) :580-585
[4]  
Dornseiffen J. W., 1979, Mededelingen van de Fakulteit Landbouwwetenschappen, Gent, V44, P867
[5]  
*EEC, 1980, 80778 EEC
[6]  
Fielding M., 1992, WATER POLLUTION RES, V27, P1
[7]   Evaluation of new polymeric sorbents with high specific surface areas using an on-line solid-phase extraction liquid chromatographic system for the trace-level determination of polar pesticides [J].
Guenu, S ;
Hennion, MC .
JOURNAL OF CHROMATOGRAPHY A, 1996, 737 (01) :15-24
[8]   SURFACE-WATER ANALYSIS (TRACE-ORGANIC CONTAMINANTS) AND EC REGULATIONS [J].
HENNION, MC ;
PICHON, V ;
BARCELO, D .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1994, 13 (09) :361-372
[9]  
HENNION MC, 1993, TECHNIQUES INSTRUMEN, V13, P4
[10]  
HIDALGO C, IN PRESS QUIM ANAL