Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial

被引:202
作者
Myers, Martin G. [1 ]
Godwin, Marshall [2 ]
Dawes, Martin [3 ]
Kiss, Alexander [4 ]
Tobe, Sheldon W.
Grant, F. Curry [5 ]
Kaczorowski, Janusz [6 ]
机构
[1] Univ Toronto, Sunnybrook Hlth Sci Ctr, Schulich Heart Ctr, Div Cardiol,Dept Med, Toronto, ON M4N 3M5, Canada
[2] Mem Univ Newfoundland, Dept Family Med, St John, NF, Canada
[3] McGill Univ, Dept Family Med, Montreal, PQ H3A 2T5, Canada
[4] Sunnybrook Hlth Sci Ctr, Dept Res Design & Biostat, Toronto, ON M4N 3M5, Canada
[5] Quinte Hlth Care, Stroke Prevent Clin, Belleville, ON, Canada
[6] Univ British Columbia, Dept Family Practice, Vancouver, BC V5Z 1M9, Canada
来源
BMJ-BRITISH MEDICAL JOURNAL | 2011年 / 342卷
关键词
MEASUREMENTS-WILL; OFFICE; DEVICE; VALIDATION; ALGORITHM; REDUCE;
D O I
10.1136/bmj.d286
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective To compare the quality and accuracy of manual office blood pressure and automated office blood pressure using the awake ambulatory blood pressure as a gold standard. Design Multi-site cluster randomised controlled trial. Setting Primary care practices in five cities in eastern Canada. Participants 555 patients with systolic hypertension and no serious comorbidities under the care of 88 primary care physicians in 67 practices in the community. Interventions Practices were randomly allocated to either ongoing use of manual office blood pressure (control group) or automated office blood pressure (intervention group) using the BpTRU device. The last routine manual office blood pressure (mm Hg) was obtained from each patient's medical record before enrolment. Office blood pressure readings were compared before and after enrolment in the intervention and control groups; all readings were also compared with the awake ambulatory blood pressure. Main outcome measure Difference in systolic blood pressure between awake ambulatory blood pressure minus automated office blood pressure and awake ambulatory blood pressure minus manual office blood pressure. Results Cluster randomisation allocated 31 practices (252 patients) to manual office blood pressure and 36 practices (303 patients) to automated office blood pressure measurement. The most recent routine manual office blood pressure (149.5 (SD 10.8)/81.4 (8.3)) was higher than automated office blood pressure (135.6 (17.3)/77.7 (10.9)) (P<0.001). In the control group, routine manual office blood pressure before enrolment (149.9 (10.7)/81.8 (8.5)) was reduced to 141.4 (14.6)/80.2 (9.5) after enrolment (P<0.001/P=0.01), but the reduction in the intervention group from manual office to automated office blood pressure was significantly greater (P<0.001/P=0.02). On the first study visit after enrolment, the estimated mean difference for the intervention group between the awake ambulatory systolic/diastolic blood pressure and automated office blood pressure (-2.3 (95% confidence interval -0.31 to -4.3)/-3.3 (-2.7 to -4.4)) was less (P=0.006/P=0.26) than the difference in the control group between the awake ambulatory blood pressure and the manual office blood pressure (-6.5 (-4.3 to -8.6)/-4.3 (-2.9 to -5.8)). Systolic/diastolic automated office blood pressure showed a stronger (P<0.001) within group correlation (r=0.34/r=0.56) with awake ambulatory blood pressure after enrolment compared with manual office blood pressure versus awake ambulatory blood pressure before enrolment (r=0.10/r=0.40); the mean difference in r was 0.24 (0.12 to 0.36)/0.16 (0.07 to 0.25)). The between group correlation comparing diastolic automated office blood pressure and awake ambulatory blood pressure (r=0.56) was stronger (P<0.001) than that for manual office blood pressure versus awake ambulatory blood pressure (r=0.30); the mean difference in r was 0.26 (0.09 to 0.41). Digit preference with readings ending in zero was substantially reduced by use of automated office blood pressure. Conclusion In compliant, otherwise healthy, primary care patients with systolic hypertension, introduction of automated office blood pressure measurement into routine primary care significantly reduced the white coat response compared with the ongoing use of manual office blood pressure measurement. The quality and accuracy of automated office blood pressure in relation to the awake ambulatory blood pressure was also significantly better when compared with manual office blood pressure.
引用
收藏
页数:9
相关论文
共 34 条
  • [1] The BpTRU automatic blood pressure monitor compared to 24 hour ambulatory blood pressure monitoring in the assessment of blood pressure in patients with hypertension
    Beckett L.
    Godwin M.
    [J]. BMC Cardiovascular Disorders, 5 (1)
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] Donner A., 2010, Design and analysis of cluster randomization trials in health research
  • [4] Pay (Adequately) For What Works: The Economic Undervaluation of Office and Ambulatory Blood Pressure Recordings
    Giles, Thomas D.
    Egan, Peter
    [J]. JOURNAL OF CLINICAL HYPERTENSION, 2008, 10 (04) : 257 - 259
  • [5] Manual and automated office measurements in relation to awake ambulatory blood pressure monitoring
    Godwin, Marshall
    Birtwhistle, Richard
    Delva, Dianne
    Lam, Miu
    Casson, Ian
    MacDonald, Susan
    Seguin, Rachelle
    [J]. FAMILY PRACTICE, 2011, 28 (01) : 110 - 117
  • [6] Definition of ambulatory blood pressure targets for diagnosis and treatment of hypertension in relation to clinic blood pressure: prospective cohort study
    Head, Geoffrey A.
    Mihailidou, Anastasia S.
    Duggan, Karen A.
    Beilin, Lawrence J.
    Berry, Narelle
    Brown, Mark A.
    Bune, Alex J.
    Cowley, Diane
    Chalmers, John P.
    Howe, Peter R. C.
    Hodgson, Jonathan
    Ludbrook, John
    Mangoni, Arduino A.
    McGrath, Barry P.
    Nelson, Mark R.
    Sharman, James E.
    Stowasser, Michael
    [J]. BMJ-BRITISH MEDICAL JOURNAL, 2010, 340 : 849
  • [7] Ishikawa J, 2009, J HYPERTENS, V27, pS4
  • [8] Mancia G, 2007, EUR HEART J, V28, P1462, DOI [10.1093/eurheartj/ehm236, 10.1093/eurheartj/eht151]
  • [9] Can sphygmomanometers designed for self-measurement of blood pressure in the home be used in office practice?
    Myers, Martin G.
    Valdivieso, Miguel
    Chessman, Mary
    Kiss, Alexander
    [J]. BLOOD PRESSURE MONITORING, 2010, 15 (06) : 300 - 304
  • [10] A proposed algorithm for diagnosing hypertension using automated office blood pressure measurement
    Myers, Martin G.
    [J]. JOURNAL OF HYPERTENSION, 2010, 28 (04) : 703 - 708