Nephron injury induced by diagnostic ultrasound imaging at high mechanical index with gas body contrast agent

被引:31
作者
Williams, Alun R.
Wiggins, Roger C.
Wharram, Bryan L.
Goyal, Meera
Dou, Chunyan
Johnson, Kent J.
Miller, Douglas L.
机构
[1] Univ Michigan, Med Ctr, Ann Arbor, MI 48109 USA
[2] Univ Michigan Hlth Syst, Dept Radiol, Ann Arbor, MI USA
[3] Univ Michigan Hlth Syst, Dept Internal Med Nephrol, Ann Arbor, MI USA
[4] Univ Michigan Hlth Syst, Dept Pathol, Ann Arbor, MI USA
关键词
diagnostic ultrasound adverse effects; bioeffects; safety; glomerular hemorrhage; ultrasound contrast agent; ultrasound perfusion imaging; flash replenishment;
D O I
10.1016/j.ultrasmedbio.2007.03.002
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The right kidney of anesthetized rats was imaged with intermittent diagnostic ultrasound (1.5 MHz; 1-s trigger interval) under exposure conditions simulating those encountered in human perfusion imaging. The rats were infused intravenously with 10 mu L/kg/min Definity (R) (Bristol-Myers Squibb Medical Imaging, Inc., N. Billerica, MA, USA) while being exposed to mechanical index (MI) values of up to 1.5 for 1 min. Suprathreshold MI values ruptured glomerular capillaries, resulting in blood filling Bowman's space and proximal convoluted tubules of many nephrons. The re-establishment of a pressure gradient after hemostasis caused the uninjured portions of the glomerular capillaries to resume the production of urinary filtrate, which washed some or all of the erythrocytes out of Bowman's space and cleared blood cells from some nephrons into urine within six hours. However, many of the injured nephrons remained plugged with tightly packed red cell casts 24 h after imaging and also showed degeneration of tubular epithelium, indicative of acute tubular necrosis. The additional damage caused by the extravasated blood amplified that caused by the original cavitating gas body. Human nephrons are virtually identical to those of the rat and so it is probable that similar glomerular capillary rupture followed by transient blockage and/or epithelial degeneration will occur after clinical exposures using similar high MI intermittent imaging with gas body contrast agents. The detection of blood in postimaging urine samples using standard hernaturia tests would confirm whether or not clinical protocols need to be developed to avoid this potential for iatrogenic injury.
引用
收藏
页码:1336 / 1344
页数:9
相关论文
共 27 条
[1]   Seeking consensus - Contrast ultrasound in radiology [J].
Albrecht, T ;
Barr, R ;
Blomley, M ;
Burns, P ;
Calliada, F ;
Campani, R ;
Correas, MCJM ;
Lafortune, M ;
Leen, E ;
Robbin, M ;
Weber, T ;
Wilson, S .
INVESTIGATIVE RADIOLOGY, 2002, 37 (04) :205-214
[2]  
Averkiou Michalakis, 2003, Ultrasound Q, V19, P27, DOI 10.1097/00013644-200303000-00004
[3]   Destruction of contrast microbubbles by ultrasound - Effects on myocardial function, coronary perfusion pressure, and microvascular integrity [J].
Ay, T ;
Havaux, X ;
Van Camp, G ;
Campanelli, B ;
Gisellu, G ;
Pasquet, A ;
Denef, JF ;
Melin, JA ;
Vanoverschelde, JLJ .
CIRCULATION, 2001, 104 (04) :461-466
[4]  
Brinker Jeffrey, 2003, Rev Cardiovasc Med, V4 Suppl 5, pS19
[5]  
CONGER JD, 1995, CLIN APPL MICROCIRCU, P239
[6]  
ELLIS EN, 1989, LAB INVEST, V60, P231
[7]  
Goldberg B.B., 2001, ULTRASOUND CONTRAST
[8]   Endothelial cell injury in venule and capillary induced by contrast ultrasonography [J].
Kobayashi, N ;
Yasu, T ;
Yamada, S ;
Kudo, N ;
Kuroki, M ;
Kawakami, M ;
Miyatake, K ;
Saito, M .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2002, 28 (07) :949-956
[9]   Progressive renal failure - Inability of podocytes to replicate and the consequences for development of glomerulosclerosis [J].
Kriz, W .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1996, 11 (09) :1738-1742
[10]   Impact of myocardial contrast echocardiography on vascular permeability: Comparison of three different contrast agents [J].
Li, P ;
Armstrong, WF ;
Miller, DL .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2004, 30 (01) :83-91