Influence of atmospheric nitric oxide concentration on the measurement of nitric oxide in exhaled air

被引:25
作者
Corradi, M
Pelizzoni, A
Majori, M
Cuomo, A
de' Munari, E
Pesci, A
机构
[1] Univ Parma, Ist Clin Malattie Apparato Resp, I-43100 Parma, Italy
[2] Azienda Reg Prevenz & Ambiente, I-43100 Parma, Italy
关键词
nitric oxide; atmospheric nitric oxide; exhaled nitric oxide;
D O I
10.1136/thx.53.8.673
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background-Measurement of nitric oxide (NO) in exhaled air shows promise as a non-invasive method of detecting lung inflammation. However, variable concentrations of NO are measured in environmental air. The aim of this study was to verify a possible relationship between exhaled NO and atmospheric NO values during high atmospheric NO days. Method-Exhaled air from 78 healthy non-smokers of mean age 35.3 years was examined for the presence of NO using a chemiluminescence NO analyser and NO levels were expressed as part per billion (ppb). The exhaled air from all the subjects was collected into a single bag and into two sequential bags. Before each test atmospheric NO was measured. Results-The mean (SE) concentration of exhaled NO collected into the single bag was 17.1 (0.6) ppb while the mean values of exhaled NO in bags 1 and 2 were 16.7 (1.3) ppb and 13.8 (1.2) ppb, respectively. The atmospheric NO concentrations registered before each test varied from 0.4 to 71 ppb. There was a significant correlation between exhaled NO in the single bag and atmospheric NO (r = 0.38, p = 0.001). The atmospheric NO concentration also correlated with exhaled NO both in bag 1 (r 0.44, p = 0.0001) and in bag 2 (r = 0.42, p 0.0001). These correlations disappeared with atmospheric NO concentrations lower than 35 ppb. Conclusions-These results indicate a relationship between atmospheric NO and NO levels measured in exhaled air, therefore exhaled NO should not be measured on very high atmospheric NO days.
引用
收藏
页码:673 / 676
页数:4
相关论文
共 24 条
  • [1] NITRIC-OXIDE AND LUNG-DISEASE
    BARNES, PJ
    BELVISI, MG
    [J]. THORAX, 1993, 48 (10) : 1034 - 1043
  • [2] Exhaled nitric oxide: A new lung function test
    Barnes, PJ
    Kharitonov, SA
    [J]. THORAX, 1996, 51 (03) : 233 - 237
  • [3] MEASUREMENT OF EXHALED NITRIC-OXIDE IN MAN
    BORLAND, C
    COX, Y
    HIGENBOTTAM, T
    [J]. THORAX, 1993, 48 (11) : 1160 - 1162
  • [4] BORLAND CDR, 1989, EUR RESPIR J, V2, P56
  • [5] Is nitric oxide in exhaled air produced at airway or alveolar level?
    Byrnes, CA
    Dinarevic, S
    Busst, C
    Bush, A
    Shinebourne, EA
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1997, 10 (05) : 1021 - 1025
  • [6] AIR-POLLUTION IN THE 1990S - CAUSE OF INCREASED RESPIRATORY-DISEASE
    DEVALIA, JL
    RUSZNAK, C
    DAVIES, RJ
    [J]. RESPIRATORY MEDICINE, 1994, 88 (04) : 241 - 244
  • [7] Airway nitric oxide in asthmatic children and patients with cystic fibrosis
    Dotsch, J
    Demirakca, S
    Terbrack, HG
    Huls, G
    Rascher, W
    Kuhl, PG
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1996, 9 (12) : 2537 - 2540
  • [8] Effect of inhaled nitric oxide or inhibition of endogenous nitric oxide formation on hyperoxic lung injury
    Garat, C
    Jayr, C
    Eddahibi, S
    Laffon, M
    Meignan, M
    Adnot, S
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1997, 155 (06) : 1957 - 1964
  • [9] ENDOGENOUS NITRIC-OXIDE IS PRESENT IN THE EXHALED AIR OF RABBITS, GUINEA-PIGS AND HUMANS
    GUSTAFSSON, LE
    LEONE, AM
    PERSSON, MG
    WIKLUND, NP
    MONCADA, S
    [J]. BIOCHEMICAL AND BIOPHYSICAL RESEARCH COMMUNICATIONS, 1991, 181 (02) : 852 - 857
  • [10] Exhaled and nasal nitric oxide measurements: Recommendations
    Kharitonov, S
    Alving, K
    Barnes, PJ
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1997, 10 (07) : 1683 - 1693