The Monte Carlo approach MCPUT for correcting pile-up distorted pulse-height spectra

被引:29
作者
Guo, WJ
Lee, SH
Gardner, RP [1 ]
机构
[1] N Carolina State Univ, CEAR, Raleigh, NC 27695 USA
[2] Hanyang Univ, iTRS, Innovat Technol Ctr Radiat Safety, Seoul 133791, South Korea
基金
美国国家卫生研究院;
关键词
Monte Carlo; MCPUT; pulse pile up; spectroscopy;
D O I
10.1016/j.nima.2004.05.089
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 [仪器科学与技术]; 080401 [精密仪器及机械]; 081102 [检测技术与自动化装置];
摘要
Pulse pile-up distortion is a common problem for radiation spectroscopy measurements involving high counting rates. The Monte Carlo pile-up to true approach (MCPUT) is proposed and benchmarked in this article for correcting pile-up distorted pulse-height spectra to true spectra. In previous work, a Monte Carlo approach was used for predicting the pile-up distorted pulse-height spectra for high counting-rate measurements ("the forward calculation"). The present work improves the previous simulation by employing a better ADC dead-time model. Based on this improved "forward calculation", the MCPUT approach introduces an iterative procedure for correcting pile-up distortions. Experiments with an Fe-55 source and a Si(Li) detector are used for benchmarking purposes. The MCPUT corrected spectrum for the high counting-rate measurement shows excellent agreement with the measured true spectrum at low counting rate with reduced chi-square as the quantitative measure. The approach is also efficient, as accurate calculations are possible in a few minutes. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:520 / 529
页数:10
相关论文
共 16 条
[1]
BALDINGER E, 1956, ADV ELECTRONICS ELEC, P255
[2]
BRISTOW Q, 1991, NUCL GEOPHYS, V5, P141
[3]
Gardner R., 1999, ADV XRAY ANAL, V41, P941
[4]
GENERALIZED METHOD FOR CORRECTING PULSE-HEIGHT SPECTRA FOR PEAK PILE-UP EFFECT DUE TO DOUBLE SUM PULSES .1. PREDICTING SPECTRAL DISTORTION FOR ARBITRARY PULSE SHAPES [J].
GARDNER, RP ;
WIELOPOLSKI, L .
NUCLEAR INSTRUMENTS & METHODS, 1977, 140 (02) :289-296
[5]
GARDNER RP, 2004, PSR528 RAD SAF INF C
[6]
GEDCKE DA, 2001, 42 INMM ANN M I NUCL
[7]
BALLISTIC DEFICIT CORRECTION IN SEMICONDUCTOR DETECTOR SPECTROMETERS [J].
GOULDING, FS ;
LANDIS, DA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :119-124
[8]
AUTOMATIC DEAD-TIME CORRECTION FOR MULTICHANNEL PULSE-HEIGHT ANALYZERS AT VARIABLE COUNTING RATES [J].
HARMS, J .
NUCLEAR INSTRUMENTS & METHODS, 1967, 53 (02) :192-&
[9]
A PRACTICAL APPROACH TO BALLISTIC DEFICIT CORRECTION [J].
HINSHAW, SM ;
LANDIS, DA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :374-377
[10]
Jenkins R., 1981, QUANTITATIVE XRAY SP