Two cases of fractured esophageal nitinol stents

被引:41
作者
Schoefl, R
Winkelbauer, F
Haefner, M
Poetzi, R
Gangl, A
Lammer, J
机构
[1] UNIV VIENNA,DEPT GASTROENTEROL & HEPATOL,A-1090 VIENNA,AUSTRIA
[2] UNIV VIENNA,DEPT INTERVENT RADIOL,A-1090 VIENNA,AUSTRIA
关键词
D O I
10.1055/s-2007-1005534
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
We report here on two cases of fractured nitinol stents in the esophagus. In case 1, the correctly inserted stent broke spontaneously shortly after insertion. In a second case, a nitinol stent broke after laser application due to tumor ingrowth with massive bleeding, In both cases, a second stent was implanted in order to reestablish food passage. The fracture of the stent in case 1 seemed to be caused by defective material, whilst in case 2 the stent broke because of thermal overstrain during laser application, The use of electrocoagulation or laser in the stent area should therefore be avoided; argon plasma coagulation may offer an effective alternative in treating tumor ingrowth, As there was a risk from piercing broken filaments with the second stent, covering a fracture using stents with tight walls or plastic tubes seems to be a more effective approach than the inserting an uncovered stent type.
引用
收藏
页码:518 / 520
页数:3
相关论文
共 20 条
[1]   SELF-EXPANDING METAL STENTS FOR PALLIATION OF MALIGNANT ESOPHAGEAL OBSTRUCTION - A PILOT-STUDY OF 8 PATIENTS [J].
BETHGE, N ;
KNYRIM, K ;
WAGNER, HJ ;
STARCK, E ;
PAUSCH, J ;
KLEIST, DV .
ENDOSCOPY, 1992, 24 (05) :411-415
[2]   MALIGNANT ESOPHAGEAL STRICTURES - TREATMENT WITH A SELF-EXPANDING NITINOL STENT [J].
CWIKIEL, W ;
STRIDBECK, H ;
TRANBERG, KG ;
VONHOLSTEIN, CS ;
HAMBRAEUS, G ;
LILLOGIL, R ;
WILLEN, R .
RADIOLOGY, 1993, 187 (03) :661-665
[3]   STENT FRACTURE IN MALIGNANT BILIARY OBSTRUCTION [J].
DONAHUE, DG ;
SALTZMAN, JR ;
KRIMS, P .
GASTROINTESTINAL ENDOSCOPY, 1993, 39 (06) :864-865
[4]   BROKEN BILIARY METAL STENT AFTER REPEATED ELECTROCOAGULATION FOR TUMOR INGROWTH [J].
ELL, C ;
FLEIG, WE ;
HOCHBERGER, J .
GASTROINTESTINAL ENDOSCOPY, 1992, 38 (02) :197-198
[5]   A NEW SELF-EXPANDABLE, NICKEL-TITANIUM COIL STENT FOR ESOPHAGEAL OBSTRUCTION - A PRELIMINARY-REPORT [J].
GOLDIN, E ;
BEYAR, M ;
SAFRA, T ;
GLOBERMAN, O ;
CRACIUN, I ;
WENGROWER, D ;
FICH, A .
GASTROINTESTINAL ENDOSCOPY, 1994, 40 (01) :64-68
[6]   HIGHLY FLEXIBLE SELF-EXPANDING MESHED METAL STENTS FOR PALLIATION OF MALIGNANT ESOPHAGOGASTRIC OBSTRUCTION [J].
GRUND, KE ;
STOREK, D ;
BECKER, HD .
ENDOSCOPY, 1995, 27 (07) :486-494
[7]  
Kozarek R A, 1994, Gastrointest Endosc Clin N Am, V4, P279
[8]   METALLIC SELF-EXPANDING STENT APPLICATION IN THE UPPER GASTROINTESTINAL-TRACT - CAVEATS AND CONCERNS [J].
KOZAREK, RA ;
BALL, TJ ;
PATTERSON, DJ .
GASTROINTESTINAL ENDOSCOPY, 1992, 38 (01) :1-6
[9]   FRACTURE OF BILIARY ENDOPROSTHESIS AFTER ENDOSCOPIC DRAINAGE FOR MALIGNANT BILIARY OBSTRUCTION - REPORT OF 2 CASES [J].
MALLAT, A ;
GIRARDIN, MFS ;
MEDURI, B ;
LIGUORY, C ;
DHUMEAUX, D .
ENDOSCOPY, 1986, 18 (06) :243-244
[10]   THE NITINOL STENT AS A PALLIATIVE MEASURE FOR INOPERABLE CARCINOMAS OF THE ESOPHAGUS AND THE CARDIA - SCOPE AND LIMITATIONS OF THE METHOD [J].
NAGELE, M ;
KONIG, C ;
TEXTOR, J ;
DECKER, P ;
HAMANN, M ;
BORN, M ;
GORICH, J ;
STEUDEL, A .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1994, 161 (02) :120-125