SAR and B1 field distributions in a heterogeneous human head model within a birdcage coil

被引:230
作者
Collins, CM
Li, SZ
Smith, MB
机构
[1] Penn State Univ, Milton S Hershey Med Ctr, Coll Med, Dept Radiol, Hershey, PA 17033 USA
[2] Penn State Univ, Milton S Hershey Med Ctr, Coll Med, Dept Cellular & Mol Physiol, Hershey, PA 17033 USA
[3] Univ Penn, Dept Bioengn, Philadelphia, PA 19104 USA
关键词
SAR; B-1; field; numerical calculations;
D O I
10.1002/mrm.1910400610
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Calculations of radiofrequency magnetic (B-1) field and specific energy absorption rate (SAR) distributions in a sphere of tissue and a multi-tissue human head model in a 12-element birdcage coil are presented. The coil model is driven in linear and quadrature modes at 63, 175, 200, and 300 MHz. Plots of B-1 field magnitude and SAR distributions, average SAR, maximum local SAR, and measures of B-1 field homogeneity and signal-to-noise ratio are given. SAR levels for arbitrary pulse sequences can be estimated from the calculated data. Maximum local SAR levels are lower at lower frequencies, in quadrature rather than in linear coils, and in linear fields oriented posterior-to-anterior rather than left-to-right in the head. It should be possible to perform many experiments in the head at frequencies up to 300 MHz without exceeding standard limits for local or average SAR levels.
引用
收藏
页码:847 / 856
页数:10
相关论文
共 27 条
  • [1] [Anonymous], 1997, Compilation of the Dielectric Properties of Body Tissues at RF and Microwave Frequencies
  • [2] RF MAGNETIC-FIELD PENETRATION, PHASE-SHIFT AND POWER DISSIPATION IN BIOLOGICAL TISSUE - IMPLICATIONS FOR NMR IMAGING
    BOTTOMLEY, PA
    ANDREW, ER
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (04) : 630 - 643
  • [3] ESTIMATING RADIOFREQUENCY POWER DEPOSITION IN BODY NMR IMAGING
    BOTTOMLEY, PA
    REDINGTON, RW
    EDELSTEIN, WA
    SCHENCK, JF
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1985, 2 (04) : 336 - 349
  • [4] BOTTOMLEY PA, 1992, ANN NY ACAD SCI, V649, P145
  • [5] INCREASED RF POWER ABSORPTION IN MR IMAGING DUE TO RF COUPLING BETWEEN BODY COIL AND SURFACE COIL
    BUCHLI, R
    SANER, M
    MEIER, D
    BOSKAMP, EB
    BOESIGER, P
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1989, 9 (01) : 105 - 112
  • [6] THE FIELD-DEPENDENCE OF NMR IMAGING .1. LABORATORY ASSESSMENT OF SIGNAL-TO-NOISE RATIO AND POWER DEPOSITION
    CHEN, CN
    SANK, VJ
    COHEN, SM
    HOULT, DI
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (05) : 722 - 729
  • [7] A method for accurate calculation of B-1 fields in three dimensions. Effects of shield geometry on field strength and homogeneity in the birdcage coil
    Collins, CM
    Li, SZ
    Yang, QX
    Smith, MB
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1997, 125 (02) : 233 - 241
  • [8] COLLINS CM, 1996, P 4 ANN M ISMRM, P253
  • [9] Durney C. H, 1985, IEEE T MICROW THEORY, V4th
  • [10] COMPARISON OF LINEAR AND CIRCULAR-POLARIZATION FOR MAGNETIC-RESONANCE IMAGING
    GLOVER, GH
    HAYES, CE
    PELC, NJ
    EDELSTEIN, WA
    MUELLER, OM
    HART, HR
    HARDY, CJ
    ODONNELL, M
    BARBER, WD
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1985, 64 (02) : 255 - 270