Detecting failed WBC-reduction processes: computer simulations of intermittent and continuous process failure

被引:9
作者
Adams, MR
Fisher, DM
Dumont, LJ
Dzik, WH
Heaton, WA
机构
[1] Univ Calif San Francisco, Dept Med Nephrol, San Francisco, CA 94143 USA
[2] Blood Ctr Pacific, San Francisco, CA USA
[3] COBE BCT Inc, Lakewood, CO USA
[4] Massachusetts Gen Hosp, Boston, MA 02114 USA
关键词
D O I
10.1046/j.1537-2995.2000.40121427.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: By regulation, ongoing process control of WBC-reduced processes is performed on 1 percent of WBC-reduced components, typically four to five samples per month. However, prospective study of the power of this small sample has been difficult. Using computer-generated "residual WBC" distributions, sample size sensitivity to continuous or intermittent WBC-reduction failure was examined. STUDY DESIGN AND METHODS: Populations of log-normally distributed values (mean +/- SD, 4.5 +/- 0.5; n = 10(5)) were generated. Continuous failure (log-normality maintained) was simulated by incrementally increasing the population mean or its SD. Intermittent failure (bimodal distributions with discrete subpopulations of WBCs > the FDA cutoff) was simulated by admiring increasing percentages of secondary outlier populations. Sample sizes of 4 to 60 were examined (500 repetitions each) for their power to detect drift or failure by standard control criteria. RESULTS: Normally distributed low variance failure was easily detected by comparison of the mean of four samples to an upper control limit (95% confidence of detecting 2% failure). However, 40 samples were required to detect > 5 percent intermittent (bimodal) failure or high variance failure with 90-percent confidence, and only if individual WBC values were compared to cutoff. CONCLUSION: Sampling error limits the detection of high variance or bimodal distributions. While the mean of a small sample is highly sensitive to shifts in a low-variance normal distribution, the detection of a high-variance bimodal population requires a large number of individual values compared to cutoff. Therefore, the number of samples required for confident failure detection depends on both the nature of the underlying distribution and the interpretive criteria. Further research is necessary to determine the true distributions of WBC-reduction process failure, as well as clinically relevant quality limits.
引用
收藏
页码:1427 / 1433
页数:7
相关论文
共 8 条
[1]  
*CTR DRUGS BIOL CT, 1995, HFN320 FDA CTR DRUGS, P17
[2]   Practical guidelines for process validation and process control of white cell-reduced blood components: Report of the biomedical excellence for safer transfusion (BEST) working party of the international society of blood transfusion (ISBT) [J].
Dumont, LJ ;
Dzik, WH ;
Rebulla, P ;
Brandwein, H .
TRANSFUSION, 1996, 36 (01) :11-20
[3]  
GRANT E, 1980, WHY CONTROL CHART WO, P85
[4]   WHITE CELL-REDUCTION IN PLATELET CONCENTRATES AND PACKED RED-CELLS BY FILTRATION - A MULTICENTER CLINICAL-TRIAL [J].
KAO, KJ ;
MICKEL, M ;
BRAINE, HG ;
DAVIS, K ;
ENRIGHT, H ;
GERNSHEIMER, T ;
GILLESPIE, MJ ;
KICKLER, TS ;
LEE, EJ ;
MCCULLOUGH, JJ ;
MCFARLAND, JG ;
NEMO, GJ ;
NOYES, WD ;
SCHIFFER, CA ;
SELL, K ;
SLICHTER, SJ ;
WOODSON, RD .
TRANSFUSION, 1995, 35 (01) :13-19
[5]   THE USE OF CONTROL CHARTS IN THE CLINICAL LABORATORY [J].
LEVEY, S ;
JENNINGS, ER .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1950, 20 (11) :1059-1066
[6]  
McCall M. J., 1997, Transfusion (Bethesda), V37, p78S
[7]  
WESTGARD JO, 1981, CLIN CHEM, V27, P493
[8]  
ZAR J, 1974, BIOSTAT ANAL, P97