Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber:: Beyond CTDI100

被引:72
作者
Dixon, Robert L. [1 ]
Ballard, Adam C. [1 ]
机构
[1] Wake Forest Univ, Sch Med, Winston Salem, NC 27157 USA
关键词
computed tomography (CT); CT dosimetry; CTDI;
D O I
10.1118/1.2757084
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This article is an experimental demonstration and authentication of a new method of computed tomography dosimetry [R. L. Dixon, Med. Phys. 30, 1272-1280 (2003)], which utilizes a short, conventional ion chamber rather than a pencil chamber, and which is more versatile than the latter. The value of CTDI100 correctly predicts the accumulated dose only for a total scan length L equal to 100 mm and underestimates the limiting equilibrium dose approached for longer, clinically relevant body scan lengths [R. L. Dixon, Med. Phys. 30, 1272-1280 (2003); K. D. Nakonechny, B. G. Fallone, and S. Rathee, Med. Phys. 32, 98-109 (2005); S. Mori, M. Endo, K. Nishizawa, T. Tsunoo, T. Aoyama, H. Fujiwara, and K. Murase, Med. Phys. 32, 1061-1069 (2005); R. L. Dixon, M. T. Munley, and E. Bayram, Med. Phys. 32, 3712-3728 (2005); R. L. Dixon, Med. Phys. 33, 3973-3976 (2006)]. Dixon [Med. Phys. 30, 1272-1280 (2003)] originally proposed an alternative using a short ion chamber and a helical scan acquisition to collect the same integral for any scan length L (and not limited 100 mm). The primary purpose of this work is to demonstrate experimentally the implementation, robustness, and versatility of this small ion chamber method in measuring the accumulated dose in the body phantom for any desired scan length L (up to the available phantom length) including the limiting equilibrium dose (symbolically CTDI infinity), and validation of the method against the pencil chamber methodology. Additionally, a simple and robust method for independently verifying the active length of a pencil chamber is described. The results of measurements made in a 400 mm long, 32 cm diameter polymethylmethacrylate body phantom using a small Farmer-type ion chamber and two pencil chambers of lengths l = 100 and 150 mm confirm that the two methodologies provide the same dose values at the corresponding scan lengths L=l. The measured equilibrium doses obtained for GE MDCT scanners at 120 kVp are CTDI infinity=1.75 CTDI100 on the central axis and 1.22 CTDI100 on the peripheral axes, illustrating a nontrivial shortfall of CTDI100 in that regard and in good agreement with comparable data [S. Mori, M. Endo, K. Nishizawa, T. Tsunoo, T. Aoyama, H. Fujiwara, and K. Murase, Med. Phys. 32, 1061-1069 (2005); J. M. Boone, Med. Phys. 34, 1364-1371 (2007)]. (C) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:3399 / 3413
页数:15
相关论文
共 25 条
[1]   New approaches to practical CT dosimetry [J].
Anderson, J ;
Chason, D ;
Arbique, G ;
Lane, T .
MEDICAL PHYSICS, 2005, 32 (06) :1907-1907
[2]   A critical look at the numerical coefficients in CTDIVOL [J].
Bakalyar, D. M. .
MEDICAL PHYSICS, 2006, 33 (06) :2003-2003
[3]   Calibration of ionization chambers in air kerma length [J].
Bochud, FO ;
Grecescu, M ;
Valley, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (09) :2477-2487
[4]   The trouble with CTDI100 [J].
Boone, John M. .
MEDICAL PHYSICS, 2007, 34 (04) :1364-1371
[5]   An improved analytical model for CT dose simulation with a new look at the theory of CT dose [J].
Dixon, RL ;
Munley, MT ;
Bayram, E .
MEDICAL PHYSICS, 2005, 32 (12) :3712-3728
[6]  
Dixon RL, 2003, MED PHYS, V30, P1272, DOI 10.1118/1.1576952
[7]   Restructuring CT dosimetry - A realistic strategy for the future requiem for the pencil chamber [J].
Dixon, Robert L. .
MEDICAL PHYSICS, 2006, 33 (10) :3973-3976
[8]   Dose and image quality for a cone-beam C-arm CT system [J].
Fahrig, Rebecca ;
Dixon, Robert ;
Payne, Thomas ;
Morin, Richard L. ;
Ganguly, Arundhuti ;
Strobel, Norbert .
MEDICAL PHYSICS, 2006, 33 (12) :4541-4550
[9]  
Intemational Electrotechnical Commission, 1997, 61674 IEC
[10]  
JENSEN A, 2006, UNPUB AAPM ANN M