Comparison of Capillary Whole Blood Versus Venous Plasma Glucose Estimations in Screening for Diabetes Mellitus in Epidemiological Studies in Developing Countries

被引:44
作者
Priya, Miranda [1 ,2 ,3 ]
Anjana, Ranjit Mohan [1 ,2 ,3 ]
Pradeepa, Rajendra [1 ,2 ,3 ]
Jayashri, Ramamurthy [1 ,2 ,3 ]
Deepa, Mohan [1 ,2 ,3 ]
Bhansali, Anil [4 ]
Mohan, Viswanathan [1 ,2 ,3 ]
机构
[1] Madras Diabet Res Fdn, Madras 600086, Tamil Nadu, India
[2] WHO Collaborating Ctr Noncommunicable Dis Prevent, Dr Mohans Diabet Special Ctr, Madras 600086, Tamil Nadu, India
[3] Int Diabet Federat Ctr Educ, Madras, Tamil Nadu, India
[4] Postgrad Inst Med Educ Res, Dept Endocrinol, Chandigarh, India
关键词
PREVALENCE; DIFFERENCE; DIAGNOSIS; ACCURACY; ANALYZER; METERS; CARE;
D O I
10.1089/dia.2010.0218
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: This study compared capillary blood glucose (CBG) measurements with venous plasma glucose (VPG) measurements in screening for diabetes and prediabetes in epidemiological studies. Methods: Four hundred seven subjects >= 20 years old (54.1% male) without previously known diabetes underwent oral glucose tolerance tests at a tertiary diabetes center in Chennai, India. Simultaneous measurements of CBG (OneTouch (R) Ultra (R) meter, LifeScan, a Johnson & Johnson Company, Milpitas, CA) and VPG (AU2700, Beckman, Fullerton, CA) were performed, both in the fasting state and 2 h after a 75-g glucose load (2-h post-glucose [PG]). Diabetes, impaired fasting glucose (IFG) and impaired glucose tolerance (IGT) were defined using American Diabetes Association (ADA) and World Health Organization (WHO) criteria. Results: The mean fasting CBG and VPG values were 122 +/- 39 mg/dL and 115 +/- 40 mg/dL, respectively, and the 2-h PG values were 203 +/- 84 mg/dL and 176 +/- 85 mg/dL, respectively. The Pearson's correlation coefficient for CBG with VPG was 0.681 (P < 0.001) in the fasting state and 0.897 (P < 0.001) for the 2-h PG load, indicating good correlation between the two methods. Based on the ADA fasting criteria, 31.9% versus 21.1% (capillary vs. venous) had diabetes, whereas based on the WHO criteria, 43.2% versus 38.6% (capillary vs. venous) had diabetes. The accuracy of identifying diabetes was 83.3% by the ADA and 90.9% by WHO criteria, for IGT it was 85.3%, and for IFG it was 66.3% by the ADA and 72.2% by the WHO criteria. Conclusion: CBG is a feasible alternative for screening of diabetes and IGT in epidemiological studies in developing countries where obtaining venous samples may be difficult.
引用
收藏
页码:586 / 591
页数:6
相关论文
共 35 条
[1]  
[Anonymous], 2006, Definition and diagnosis of diabetes mellitus and intermediate hyperglycaemia: Report of a WHO/IDF consultation
[2]  
[Anonymous], 2003, SCREEN TYP 2 DIAB RE
[3]  
*AS PAC TYP 2 DIAB, TYP 2 DIAB PRACT TAR
[4]   WHAT IS BLOOD-GLUCOSE - CAN IT BE MEASURED [J].
BURRIN, JM ;
ALBERTI, KGMM .
DIABETIC MEDICINE, 1990, 7 (03) :199-206
[5]   Measurement of blood glucose:: comparison between different types of specimens [J].
Carstensen, Bendix ;
Lindstrom, J. ;
Sundvall, J. ;
Borch-Johnsen, K. ;
Tuomilehto, J. .
ANNALS OF CLINICAL BIOCHEMISTRY, 2008, 45 :140-148
[6]   Comparability of venous and capillary glucose measurements in blood [J].
Colagiuri, S ;
Sandbæk, A ;
Carstensen, B ;
Christensen, J ;
Glumer, C ;
Lauritzen, T ;
Borch-Johnsen, K .
DIABETIC MEDICINE, 2003, 20 (11) :953-956
[7]   Glucose measurement: Confounding issues in setting targets for inpatients management [J].
Dungan, Kathleen ;
Chapman, John ;
Braithwaite, Susan S. ;
Buse, John .
DIABETES CARE, 2007, 30 (02) :403-409
[8]  
Fink Kenneth S, 2002, Diabetes Technol Ther, V4, P627, DOI 10.1089/152091502320798259
[9]   Agreement between bedside blood and plasma glucose measurement in the ICU setting [J].
Finkielman, JDW ;
Oyen, LJ ;
Afessa, B .
CHEST, 2005, 127 (05) :1749-1751
[10]   Relative accuracy of the BD Logic® and FreeStyle® blood glucose meters [J].
Fox, Larry A. ;
Beck, Roy ;
Steffes, Michael ;
Chase, Peter ;
Coffey, Julie ;
Wysocki, Tim ;
Buckingham, Bruce ;
Weinzimer, Stuart ;
Tamborlane, William ;
Kollman, Craig ;
Ruedy, Katrina .
DIABETES TECHNOLOGY & THERAPEUTICS, 2007, 9 (02) :165-168